User manual for Varian Turbo-V60 Leak Detector PCB Controller, covering safety, installation, operation, and maintenance.
This manual provides instructions for operating and maintaining the Turbo-V60 Leak Detector PCB Controller. It includes safety warnings, product description, installation steps, operational procedures, and maintenance guidelines. The controller is a microprocessor-controlled solid-state frequency converter designed to drive the Turbo-V60 pump. It features remote start/stop and output control via auxiliary connectors. The manual also includes a return request form and health/safety certification requirements for equipment returns. This manual is written in Chinese.
| Input Voltage | 120 Vac ±20% or 220 Vac ±20% |
|---|---|
| Output Voltage | 54 Vac nominal ±10%, 3-phase |
| Operating Temperature | 0°C to +40°C |
| Weight | 0.5 Kg (1.1 Ibs) |
1167 Hz, ±2%.
Use the three-pin male connector on the transformer primary: for 120 Vac connect to pin 2 (white) and pin 3 (blue); for 220 Vac connect to pin 1 (brown) and pin 3 (blue).
0°C to +40°C.