Pre-installation instructions for Varian 210-MS, 220-MS, and 225-MS GC/MS ion trap mass spectrometers covering site, environmental, power, and gas requirements.
This document provides pre-installation instructions and a checklist for Varian 210-MS, 220-MS, and 225-MS GC/MS ion trap mass spectrometers. It guides users through site preparation requirements before installation is scheduled.
It covers safety, entrance clearance, bench space and load, vibration, temperature, humidity, particulate matter, ventilation, power, gas requirements, shipping carton inspection, and installation-related steps.
| Laboratory temperature range | 16 to 30 °C (61 to 86 °F) |
|---|---|
| Relative humidity range | 20 to 80%, no condensation |
| Laboratory entrance width | at least 92 cm (36 in.) |
| Shipping carton dimensions | 72 cm (28 in.) wide by 92 cm (36 in.) long |
| Bench depth | at least 84 cm (33 in.) |
| Bench length with 431-GC | 120.4 cm (47.7 in.) |
| Bench load with 431-GC and 210-MS/220-MS | 63.9 kg (141.4 lb) |
| Bench load with 431-GC and 225-MS | 74.9 kg (165.4 lb) |
| Bench length with 450-GC | 153.4 cm (60.65 in.) |
| Bench load with 450-GC and 210-MS/220-MS | 83.9 kg (185.4 lb) |
| Bench load with 450-GC and 225-MS | 94.9 kg (209.4 lb) |
| Foreline pump vacuum line length | 1.2 m (48 in.) |
| MS vacuum line supplied | 1.7 m (66 in.) |
| Maximum bench height for pump | 91 cm (36 in.) |
| Foreline pump weight | 11 kg (24 lb) |
| Bench hole diameter for vacuum hose | 3.8 cm (1.5 in.) |
| PC distance from MS | within 3 m (10 ft) |
| 210-MS/220-MS dimensions and weight | 46 x 25 x 64 cm (18 x 10 x 25 in.); 23 kg (51 lb) |
| 225-MS dimensions and weight | 46 x 25 x 76.2 cm (18 x 10 x 30 in.); 34 kg (75 lb) |
| 431-GC dimensions and weight | 46 x 33 x 56 cm (18 x 13 x 22 in.); 23 kg (51 lb) |
| 450-GC dimensions and weight | 53 x 66 x 56 cm (21 x 26 x 22 in.); 43 kg (95 lb) |
| CI reagent gas purity | 99.99% (methane, isobutane, or ammonia) |
| Helium purity | 99.999% |
| Average steady-state heat load | 6,000 Btu |
| Short-term heat dissipation during startup | 15,000 Btu |
| Space to left of GC/MS | at least 30 cm (12 in.) |
| Space behind 210-MS/220-MS | 15 to 30 cm (6 to 12 in.) |
| Space behind 225-MS | 25 to 30 cm (10 to 12 in.) |
| Space above GC/MS | at least 76 cm (30 in.) |
The laboratory temperature must be maintained between 16 and 30 °C.
Relative humidity must be between 20 and 80% with no condensation.
Methane, isobutane, or ammonia at 99.99% purity.
Helium must be 99.999% pure.