User manual for Thermo Scientific STANDARD Series thermostats SC 100, SC 150 and SC 150L covering installation, operation, maintenance and troubleshooting.
This manual covers the Thermo Scientific STANDARD Series thermostats SC 100, SC 150 and SC 150L, which are used with refrigerated and heated baths. All models can pump to an external system and have a digital display, touch pad, five programmable setpoint temperatures, and acoustic and optical alarms. The SC 150 and SC 150L additionally offer adjustable high temperature protection and independent low liquid level protection.
The manual includes safety warnings, specifications, installation instructions (electrical requirements, tubing, fluids, filling and draining), operation of the thermostat menus and settings, accessory installation, preventive maintenance, troubleshooting error displays, and serial communications. It also contains the Declaration of Conformity, a Return Materials Authorization Form and warranty details.
| Temperature Range (SC 100) | Ambient +13°C to +100°C |
|---|---|
| Temperature Range (SC 150 / SC 150L) | Ambient +13°C to +150°C |
| Temperature Stability | ±0.02°C (all models) |
| Heater Capacity (230V/115V) | 2000/1200 Watts |
| Immersion Depth | 75 to 145 mm (SC 100/SC 150); 75 to 190 mm (SC 150L) |
| Dimensions (H x W x D) | 336 x 138 x 199 mm (SC 100/SC 150); 384 x 138 x 199 mm (SC 150L) |
| Net Weight | 3.3 kg (7.3 lb) |
| USB Interface | No (SC 100); Yes (SC 150, SC 150L) |
| Max Flow (12 mm id hose) | 12.5 lpm (3.3 gpm) |
Thermo Scientific Laboratory Temperature Control Products have a warranty against defective parts and workmanship for 36 months from date of shipment.
The SC 100 is designed to be used only with water or glycol-water.
Refrigerated units should be left in an upright position for 24 hours before starting so the lubrication oil drains back into the compressor.
Ensure the fluid is at a safe temperature of 20°C to 55°C before handling or draining.