User guide for the Thermo Scientific OrbitorRS2, covering safety, specifications, operation, teaching locations, and maintenance.
This user guide describes the Thermo Scientific OrbitorRS2 and its operation, including safety information, physical and environmental specifications, control methods, barcode reading, location teaching, and maintenance procedures.
The guide also covers cleaning, safety-control testing, fuse replacement for BenchTrak-mounted units, and disposal requirements for the OrbitorRS2.
| Height | 724 mm (28.5 in.) |
|---|---|
| Weight | 25 kg (55 lbs) |
| Vertical workspace | 16 - 591 mm (0.6 - 23.2 in.) |
| Horizontal workspace diameter | 820 mm (32.2 in.) |
| Gripper maximum payload | 320 g (11.3 oz) |
| Gripper opening distance | 75 - 130 mm (2.97 - 5.13 in.) |
| Gripper maximum closing force | 8.9 N (32 ozf) |
| Container type | SBS standard |
| Container orientation | Landscape or portrait |
| Ambient operating temperature | 15 - 40°C (59 - 104°F) |
| Storage temperature | -20 - 60°C (-4 - 140°F) |
| Humidity | 20 - 80% RH up to 31°C (88°F), decreasing linearly to 50% RH at 40°C (104°F); max 80% RH non-condensing |
| Atmosphere | Office/laboratory environment (clean, low-dust, and dry) |
| Maximum operating altitude | 2,000 m (6,560 ft) above sea level |
| Installation location | For indoor use only |
| Voltage power requirement | 100 - 240 +/- 10% VAC auto-switching |
| BenchTrak MoverLink fuse rating | 5 A/125 V FAST |
| BenchTrak MoverLink fuse size | 5 x 20 mm |
| BenchTrak MoverLink fuse manufacturer part number | Littelfuse 0235005.HXP |
It must undergo a maintenance inspection once per year, performed by a qualified Thermo Fisher Scientific technician.
No. The guide states not to autoclave any part of the Orbitor.
Only trained and authorized personnel may repair the Orbitor.
It should be cleaned at least once per month to remove dust or debris.