This manual provides installation, operation, troubleshooting, and care instructions for the Thermo Fisher Scientific Dionex IonPac CS14 analytical columns…
The IonPac CS14 analytical columns and CG14 guard columns are designed for the separation of alkyl and alkanolamines from Group I and Group II cations. The stationary phase consists of a 8.0 μm poly (ethylvinylbenzene/divinylbenzene) macroporous substrate resin functionalized with a relatively weak carboxylic acid, providing both cation exchange and reverse phase properties. These columns are compatible with eluents having a pH from 0 to 14 and can be used with various organic solvents including acetonitrile, methanol, isopropanol, and tetrahydrofuran.
The stationary phase consists of a 8.0 μm poly (ethylvinylbenzene/divinylbenzene) macroporous substrate resin functionalized with a relatively weak carboxylic acid.
Yes, the column is solvent-compatible and can be used with 0-100% acetonitrile, methanol, isopropanol, or tetrahydrofuran.
The recommended maximum back pressure for the 4-mm CS14 analytical column is less than 1,200 psi (8.27 MPa).