Siemens presentation on the SITRANS P320/P420 pressure transmitters: features, specifications, DSIII comparison, display and operation, and ordering.
This Siemens Vietnam presentation introduces the SITRANS P320 and P420 pressure transmitters as the next generation of pressure devices. It covers the pressure portfolio versions for gauge, absolute and differential pressure and level measurement with differential pressure cells, then lists the headline features: a larger HMI display with NAMUR NE107 support and a quick start wizard, remote safety handling for SIL commissioning, HART7 communication with long tag support, and reduced response time.
The document also gives a product data comparison against the DSIII (linear characteristic, temperature error, static pressure effect, total performance, long term stability, step response time and SIL assessment), describes the Trend Log data logging and limit monitoring functions, and closes with model number and ordering information plus remote seal configuration types 7MF0812, 7MF0813 and 7MF0814.
| Communication protocol | HART7 with Long tag support |
|---|---|
| Linear characteristic (P320) | 1:1…5:1: 0.065%; TD > 5:1: (0.005 x r + 0.05)% |
| Linear characteristic (P420) | 1:1…5:1: 0.04%; TD > 5:1: (0.004 x r + 0.045)% |
| Temperature error (P420) | (0.0125 x r + 0.0625)% / 28K |
| Total performance (TD 1:1) | SITRANS P320 0.18%; SITRANS P420 0.12% |
| Long term stability | 0.125% / 5a |
| Step response time | 135 ms DP / 105 ms P for SITRANS P320 and P420 |
| Safety | Developed according to IEC61508 for SIL2/3 applications, SFF > 92% |
| Housing material | Aluminum or stainless steel 316L |
| Membrane material | Stainless steel 316L, alloy C276, Monel 400, tantalum or gold |
| Trend Log capacity | 2 process values, max. 1500 points |
They use HART7, including Long tag support.
The Trend Log records 2 process values with a maximum of 1500 points.
It allows safety commissioning and safety validation to be performed remotely, for example via SIMATIC PDM.
The step response time is 135 ms for differential pressure and 105 ms for pressure.