Siemens Inveon Scanners and Inveon Acquisition Workplace 1.5 Service Pack 1 User Manual

User manual for Siemens Inveon scanners and Inveon Acquisition Workplace 1.5 with Service Pack 1, covering calibration, quality control and scanning.


Siemens Inveon Scanners and Inveon Acquisition Workplace 1.5 Service Pack 1 User Manual - cover page
Brand
Siemens
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
329
File format
PDF
File size
6.4 MB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
57D5B427010E0A469C4B22733F34D08D

About this manual

This is the Inveon User Manual from Siemens Medical Solutions USA for the Inveon scanners — the Inveon Dedicated PET Scanner (D-PET) and the Inveon Multimodality Scanner (MM) — and the Inveon Acquisition Workplace 1.5 with Service Pack 1 software. It covers the scanners, the computers and docking configurations, hardware installation, maintenance, cleaning and repair, calibration and quality control schedules, and the CT, PET and SPECT modalities.

The manual also describes Inveon Acquisition Workplace (IAW) operations, data storage, microQ and other software such as ImageJ and ASIPro, calibration to the Hounsfield scale, phantoms, detector setup, normalization, attenuation correction and transformation matrices, along with troubleshooting, a glossary and the Inveon engineering specifications.

What's inside

  • Introducing the Inveon Platform
  • Calibration and Quality Control Schedules
  • Inveon Dedicated PET Scanner (D-PET)
  • Inveon Multimodality Scanner (MM)
  • Docking the Inveon Multimodality and D-PET Scanners
  • Inveon Acquisition Workplace (IAW)
  • CT Procedures
  • PET Procedures
  • SPECT Procedures
  • Inveon Engineering Specifications

Specifications

Unit weight (multimodality scanner)1060–1270 kg (2,335–2,800 lb)
Unit height198 cm (78 in)
Unit width173 cm (68 in)
Unit depth163 cm (64.3 in)
Operating room temperature18–24 °C (65–75 °F) with ≤3 °C (5.4 °F) change per hour
Operating humidity30–70% non-condensing
Maximum heat generation1905 W (6500 Btu/h)
Bore diameter12 cm
Power requirements12 A, 120 V, 50/60 Hz dedicated circuit or 6 A, 200-240 V, 50/60 Hz dedicated circuit (internal fuse 18 A)

Frequently asked questions

What power supply does the Inveon multimodality scanner need?

It uses a dedicated 12 A, 120 V, 50/60 Hz circuit or a dedicated 6 A, 200-240 V, 50/60 Hz circuit, each with an 18 A internal fuse.

What is the bore diameter of the multimodality scanner?

The bore diameter is 12 cm.

What room conditions does the system require?

The operating room is kept at 18–24 °C (65–75 °F) with no more than 3 °C change per hour, and humidity of 30–70% non-condensing.

What does this revision of the Inveon User Manual add?

It includes an addendum of new topics and major revisions, such as Calibration and QC Schedules, Using the Bore Tunnel, Cooling the CT Camera and Data Storage.