User manual for Siemens Inveon scanners and Inveon Acquisition Workplace 1.5 with Service Pack 1, covering calibration, quality control and scanning.
This is the Inveon User Manual from Siemens Medical Solutions USA for the Inveon scanners — the Inveon Dedicated PET Scanner (D-PET) and the Inveon Multimodality Scanner (MM) — and the Inveon Acquisition Workplace 1.5 with Service Pack 1 software. It covers the scanners, the computers and docking configurations, hardware installation, maintenance, cleaning and repair, calibration and quality control schedules, and the CT, PET and SPECT modalities.
The manual also describes Inveon Acquisition Workplace (IAW) operations, data storage, microQ and other software such as ImageJ and ASIPro, calibration to the Hounsfield scale, phantoms, detector setup, normalization, attenuation correction and transformation matrices, along with troubleshooting, a glossary and the Inveon engineering specifications.
| Unit weight (multimodality scanner) | 1060–1270 kg (2,335–2,800 lb) |
|---|---|
| Unit height | 198 cm (78 in) |
| Unit width | 173 cm (68 in) |
| Unit depth | 163 cm (64.3 in) |
| Operating room temperature | 18–24 °C (65–75 °F) with ≤3 °C (5.4 °F) change per hour |
| Operating humidity | 30–70% non-condensing |
| Maximum heat generation | 1905 W (6500 Btu/h) |
| Bore diameter | 12 cm |
| Power requirements | 12 A, 120 V, 50/60 Hz dedicated circuit or 6 A, 200-240 V, 50/60 Hz dedicated circuit (internal fuse 18 A) |
It uses a dedicated 12 A, 120 V, 50/60 Hz circuit or a dedicated 6 A, 200-240 V, 50/60 Hz circuit, each with an 18 A internal fuse.
The bore diameter is 12 cm.
The operating room is kept at 18–24 °C (65–75 °F) with no more than 3 °C change per hour, and humidity of 30–70% non-condensing.
It includes an addendum of new topics and major revisions, such as Calibration and QC Schedules, Using the Bore Tunnel, Cooling the CT Camera and Data Storage.