R&S FSH3-TV TV Analyzer VSWR Measurement in 75Ω Systems Application Note

This application note details the procedure for measuring Voltage Standing Wave Ratio (VSWR) in 75Ω systems using the R&S FSH3-TV TV analyzer with 50Ω and…


R&S FSH3-TV TV Analyzer VSWR Measurement in 75Ω Systems Application Note - cover page
Brand
ROHDE&SCHWARZ
Category
Other Electronics
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
1 MB
Published
23 October, 2026
Updated
23 October, 2026
MD5 checksum
E8CCA89D795D2D9FA31E898F87E45E74

About this manual

This document describes how to measure the Voltage Standing Wave Ratio (VSWR) in 75Ω systems using the R&S FSH3-TV handheld TV analyzer, which is configured for 50Ω system impedances. It covers the theoretical basics of VSWR and reflection coefficients, explaining how impedance mismatches lead to signal reflections and losses.

The guide provides step-by-step measurement procedures for two scenarios: using a 50Ω calibration kit (R&S FSH-Z2 or FSH-Z3) with an R&S RAM matching pad, which requires a correction factor, and using a 75Ω calibration kit (R&S ZCAN), which does not require further correction. It also explains how to perform Smith chart measurements to visualize impedance matching.

What's inside

  • Overview
  • Requirements
  • Basics of VSWR
  • Measurement Procedure
  • Procedure with 50Ω Calibration Kit
  • Procedure with 75Ω Calibration Kit
  • Smith Chart Measurement
  • Reflection coefficient correction factor
  • Ordering Information

Specifications

Product TypeHandheld TV Analyzer
VSWR Bridge Frequency Range10 MHz to 3 GHz
VSWR Bridge with DC Bias Frequency Range10 MHz to 6 GHz

Frequently asked questions

Does the R&S RAM matching pad provide matching in both directions?

Yes, the R&S RAM matching pad is bidirectional, providing a match for both forward and reflected signals.

What impedance value must be selected on the Smith Chart to measure systems?

You must select a characteristic impedance (Z0) of 75 Ω in the Marker Mode settings.

What trace offset value is required when using a 50 Ω calibration kit?

You must set a trace offset of 11.5 dB to compensate for the matching pad attenuation.