RKI Instruments Eagle Portable Gas Detector Operation and Calibration Manual

Spanish instruction manual for the RKI Instruments Eagle portable gas detector, covering operation, display functions, alarms and calibration.


RKI Instruments Eagle Portable Gas Detector Operation and Calibration Manual - cover page
Brand
RKI Instruments
Category
Monitoring Equipment
Document type
User Manual
Language
Spanish
Pages
10
File format
PDF
File size
129 KB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
9966DC5751E8CEAF5166233F06F4788E

About this manual

The RKI Instruments Eagle is a portable gas detector that simultaneously measures combustible gas in %LEL or ppm, oxygen deficiency, carbon monoxide and hydrogen sulfide. This Spanish instruction manual covers operation and calibration, and states that the instrument is intrinsically safe for Class I, Division I, Groups A, B, C and D hazardous environments. It describes the control panel switches, the sample draw system and the electronic cards, which the manual says must be serviced only by RKI Instruments or authorized personnel.

The manual explains preparation, power-up diagnostics and normal operation, along with the display functions PEAK, TIME IN OPERATION and TWA/STEL. It describes latching, self-resetting, malfunction and battery alarms, then details calibration: required gas cylinders and flow regulator, zero adjustment, automatic calibration with the RKI 4-gas cylinder and single calibration. This manual is written in Spanish.

What's inside

  • Componentes y controles
  • Sistema para atraer la muestra
  • Operación
  • Funciones en el presentador
  • Reinicializando las alarmas
  • Alarmas de enclavamiento
  • Alarmas de funcionamiento defectuoso
  • Alarma de batería
  • Calibración
  • Calibración automática

Specifications

Gases detectedCombustible gas (%LEL or ppm), oxygen deficiency (O2), carbon monoxide (CO), hydrogen sulfide (H2S)
Intrinsic safety ratingClass I, Division I, Groups A, B, C and D
Sample draw distanceUp to 75 ft with proper configuration
Minimum battery voltage4.5 V
Oxygen zero setting20.9% (ambient oxygen level)

Frequently asked questions

What gases does the Eagle detector measure?

It simultaneously detects combustible gas in %LEL or ppm, oxygen deficiency, carbon monoxide and hydrogen sulfide.

How do I turn the Eagle detector off?

Press and hold POWER/ENTER until a series of beeps sound and the display shows the message "GOOD-BYE"; wait for the message to disappear before turning the instrument on again.

What does the TWA/STEL function show?

TWA readings show average toxic sensor concentrations over the last 8 hours, while STEL readings show average concentrations over the last 15 minutes.

What does the battery alarm display?

When the voltage reaches the permitted limit, the display shows BATTERY MIN.4.5V CHANGE BATTERY, and the Eagle cannot be used to monitor gas concentrations.