RKI Instruments M2 Transmitter Operator's Manual

This manual provides operating instructions, configuration, maintenance, and troubleshooting for the RKI Instruments M2 fixed-mount gas transmitter.


RKI Instruments M2 Transmitter Operator's Manual - cover page
Brand
RKI Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
72
File format
PDF
File size
1.2 MB
Published
12 November, 2026
Updated
12 November, 2026
MD5 checksum
190C498C0880F5E08AAFD776D82E5385

About this manual

The RKI Instruments M2 Transmitter is a fixed-mount, continuous-monitoring detector head designed for safety applications. It features an LCD display visible through a cover window and provides a 4-20 mA signal output for integration with controllers or recording devices. The device supports non-intrusive calibration using a magnetic wand accessory and includes three sets of relay contacts for alarm control.

What's inside

  • Introduction
  • Description
  • Installation & Startup
  • Operation
  • Configuration Mode
  • Gas Type Mode
  • Maintenance
  • RS-485 Modbus Output

Specifications

Input Power19 - 30 VDC
Weight4.5 lbs
Signal Output4 to 20 mA, 500 ohms impedance max
Operating Temperature (All Others)-20°C to 50°C
Accuracy± 5% of full scale
Relay Rating115 VAC 5 amps

Frequently asked questions

How often should the instrument be calibrated?

Typical calibration frequencies are between 3 and 6 months, but this depends on usage and sensor types.

Can calibration be performed without opening the junction box?

Yes, calibration can be performed non-intrusively using a magnetic wand accessory.

What is the maximum impedance for the 4-20 mA signal output?

The maximum impedance for the signal output is 500 ohms.