RKI Instruments M2A Transmitter Operator's Manual

This manual introduces the M2A transmitter user manual and warranty information


RKI Instruments M2A Transmitter Operator's Manual - cover page
Brand
RKI Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
80
File format
PDF
File size
865 KB
Published
15 October, 2023
Updated
15 October, 2023
MD5 checksum
517EE890C61B65E7B605AB9894A1A5C4

About this manual

The M2A Transmitter is a fixed-mount, continuous-monitoring detector head designed for detecting combustible gases, carbon dioxide, oxygen, and various toxic gases. It features an OLED display, push-button controls, and supports non-intrusive calibration via a magnetic wand. The device provides a 4-20 mA signal output and an RS-485 Modbus output for integration with monitoring controllers and programmable logic controllers.

What's inside

  • Introduction
  • Description
  • Installation & Startup
  • Operation
  • Configuration Mode
  • Gas Type Mode
  • Maintenance
  • RS-485 Modbus Output

Specifications

Target Gas/Detection Range (Combustible Gas)0 - 100% LEL, 1% LEL increments
Target Gas/Detection Range (Oxygen)0 - 25.0 % volume, 0.1% increments
Target Gas/Detection Range (Carbon Monoxide)0 - 300 ppm, 1 ppm increments
Input Power10 - 30 VDC
Weight4.5 lbs.
Signal Output4 to 20 mA
Operating Temperature (Catalytic LEL/ppm)-40°C to 75°C
Accuracy (Catalytic Combustible Gas)± 5% of reading or ± 2% LEL

Frequently asked questions

How often should the M2A be calibrated?

Typical calibration frequencies are between 3 and 6 months, though this depends on usage and sensor types.

Can the M2A be calibrated without opening the device?

Yes, calibration can be performed non-intrusively using a magnetic wand accessory.

What is the warranty period for the M2A?

RKI Instruments warrants the equipment to be free from defects for one year from the date of shipment.