Product specification for the PANDA P2 intraoral digital impression scanner, covering components, dimensions, software requirements, sterilization and care.
The PANDA P2 is an intraoral digital impression instrument manufactured by Ziyang Freqty Medical Equipment Co., Ltd. It uses an optical scanning method to obtain three-dimensional shape data of the surfaces of teeth, gums and other tissue, and outputs digital impression data that can be used in CAD/CAM denture design and processing. The product is composed of the probe (probe body and P2 probe head assembly), power adapter, calibrator, supporting software and probe bracket.
The document lists the main dimensions (85 mm length into the mouth, 15 mm depth of field, 216 x 40 x 36 mm total size), scanning accuracy of 15 um or better and repeatability precision of 10 um or better, along with operating conditions of 5 to 40 degrees C and 760-1060 hPa, and AC 100-240 V 50/60 Hz 30 VA input power. It also covers the 450 nm and 520 nm visible laser with maximum power below 0.4 mW, computer requirements for the scan software, cleaning and steam sterilization of the P2 probe head (121 degrees C for 15 min or 134 degrees C for 6 min, no more than 30 times), transport and storage conditions, a five-year lifespan, the parts list and EMC information.
| Length into the mouth | 85 mm |
|---|---|
| Depth of field | 15 mm |
| Total size | 216 mm (L) x 40 mm (W) x 36 mm (H) |
| Scanning depth | 0~15 mm |
| Scanning accuracy / repeatability precision | 15 um or better / 10 um or better |
| Operating temperature | 5 to 40 degrees C |
| Atmospheric pressure | 760 hPa ~ 1060 hPa |
| Rated input power | AC 100-240 V, 50/60 Hz, 30 VA |
| Laser wavelength | 450 nm and 520 nm |
| Laser maximum power | less than 0.4 mW |
| Software CPU requirement | Intel i7-12700H / Intel i7-11800H or above |
| Software RAM / storage requirement | 16G/32G RAM, SSD 512G or above |
| Software GPU requirement | RTX2060/RTX3060 or above |
| Operating system | Windows 10/11 64 bit |
| Lifespan | Five years |
| Recommended sterilization of P2 probe head | Moist heat steam, 121 degrees C for 15 min or 134 degrees C for 6 min |
| Repeated sterilization limit | Not more than 30 times |
It is recommended to use moist heat steam sterilization at 121 degrees C for 15 min or at 134 degrees C for 6 min.
It is recommended that the number of repeated sterilizations is not more than 30 times; the head should be replaced if its appearance is damaged or after 30 sterilizations.
An Intel i7-12700H or i7-11800H CPU (or above), 16G/32G RAM, SSD 512G or above, an RTX2060/RTX3060 or above graphics card and Windows 10/11 64 bit.
It uses a visible laser with wavelengths of 450 nm and 520 nm, a beam divergence angle of 33 degrees, pulse width 25 ms, repetition frequency 30 Hz and maximum power less than 0.4 mW.