PANDA P2 Intraoral Digital Impression Scanner Product Specification

Product specification for the PANDA P2 intraoral digital impression scanner, covering components, dimensions, software requirements, sterilization and care.


PANDA P2 Intraoral Digital Impression Scanner Product Specification - cover page
Brand
Panda
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
16
File format
PDF
File size
5.6 MB
Published
10 November, 2026
Updated
10 November, 2026
MD5 checksum
1129D8D7894B318B6861A7F6E1A22F6D

About this manual

The PANDA P2 is an intraoral digital impression instrument manufactured by Ziyang Freqty Medical Equipment Co., Ltd. It uses an optical scanning method to obtain three-dimensional shape data of the surfaces of teeth, gums and other tissue, and outputs digital impression data that can be used in CAD/CAM denture design and processing. The product is composed of the probe (probe body and P2 probe head assembly), power adapter, calibrator, supporting software and probe bracket.

The document lists the main dimensions (85 mm length into the mouth, 15 mm depth of field, 216 x 40 x 36 mm total size), scanning accuracy of 15 um or better and repeatability precision of 10 um or better, along with operating conditions of 5 to 40 degrees C and 760-1060 hPa, and AC 100-240 V 50/60 Hz 30 VA input power. It also covers the 450 nm and 520 nm visible laser with maximum power below 0.4 mW, computer requirements for the scan software, cleaning and steam sterilization of the P2 probe head (121 degrees C for 15 min or 134 degrees C for 6 min, no more than 30 times), transport and storage conditions, a five-year lifespan, the parts list and EMC information.

What's inside

  • Product Composition
  • Main Performance and Main Dimension Specifications
  • Product Function and Intended Use
  • Environmental & Working Power Requirements
  • Product Hardware Installation Instructions
  • Software Operation Configuration Requirements
  • Care and Maintenance Methods
  • Transport and Storage Life
  • Parts List
  • About EMC

Specifications

Length into the mouth85 mm
Depth of field15 mm
Total size216 mm (L) x 40 mm (W) x 36 mm (H)
Scanning depth0~15 mm
Scanning accuracy / repeatability precision15 um or better / 10 um or better
Operating temperature5 to 40 degrees C
Atmospheric pressure760 hPa ~ 1060 hPa
Rated input powerAC 100-240 V, 50/60 Hz, 30 VA
Laser wavelength450 nm and 520 nm
Laser maximum powerless than 0.4 mW
Software CPU requirementIntel i7-12700H / Intel i7-11800H or above
Software RAM / storage requirement16G/32G RAM, SSD 512G or above
Software GPU requirementRTX2060/RTX3060 or above
Operating systemWindows 10/11 64 bit
LifespanFive years
Recommended sterilization of P2 probe headMoist heat steam, 121 degrees C for 15 min or 134 degrees C for 6 min
Repeated sterilization limitNot more than 30 times

Frequently asked questions

How should the P2 probe head assembly be sterilized?

It is recommended to use moist heat steam sterilization at 121 degrees C for 15 min or at 134 degrees C for 6 min.

How many times can the P2 probe head be sterilized?

It is recommended that the number of repeated sterilizations is not more than 30 times; the head should be replaced if its appearance is damaged or after 30 sterilizations.

What computer configuration does the scan software require?

An Intel i7-12700H or i7-11800H CPU (or above), 16G/32G RAM, SSD 512G or above, an RTX2060/RTX3060 or above graphics card and Windows 10/11 64 bit.

What laser does the PANDA P2 use?

It uses a visible laser with wavelengths of 450 nm and 520 nm, a beam divergence angle of 33 degrees, pulse width 25 ms, repetition frequency 30 Hz and maximum power less than 0.4 mW.