Nellcor PM100N Bedside SpO2 Patient Monitoring System Product Leaflet

Product leaflet for the Nellcor PM100N Bedside SpO2 Patient Monitoring System with features and specifications.


Nellcor PM100N Bedside SpO2 Patient Monitoring System Product Leaflet - cover page
Brand
Nellcor
Category
Professional medical equipment
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
585 KB
Published
13 November, 2026
Updated
13 November, 2026
Model
PM100N
MD5 checksum
206A462214298BE28A81508A150D0076

About this manual

This document is a two-page product leaflet for the Nellcor PM100N Bedside SpO2 Patient Monitoring System. It describes pulse oximetry monitoring features including SatSeconds alarm management, Sleep Study mode, Homecare mode, and adult, pediatric, and neonate modes.

The leaflet also lists performance, electrical, environmental, physical, compliance, classification, output, display, alarm, accessory, mode, and connectivity specifications for the monitor.

What's inside

  • Features and specifications
  • Performance
  • Electrical
  • Environmental
  • Physical Characteristics
  • Equipment Compliance Standards
  • Equipment Classifications
  • Available Modes
  • Connectivity
  • Simple set up and maintenance

Specifications

SpO2 measurement range1% to 100%
Pulse rate measurement range20 to 250 bpm
Pulse amplitude measurement range0.03% to 20%
SpO2 accuracy (adult, 70% to 100%)±2 digits
SpO2 accuracy (adult and neonate low saturation, 60% to 80%)±3 digits
SpO2 accuracy (neonate, 70% to 100%)±2 digits
SpO2 accuracy (low perfusion, 70% to 100%)±2 digits
SpO2 accuracy (adult and neonate with motion, 70% to 100%)±4 digits
Pulse rate accuracy (adult and neonate, 20 to 250 bpm)±3 digits
Pulse rate accuracy (low perfusion, 20 to 250 bpm)±3 digits
Pulse rate accuracy (adult and neonate with motion, 20 to 250 bpm)±5 digits
Power requirements100 to 240 VAC, 50/60 Hz, 45 VA
Fuse ratingFast-acting 2 A 32VAC/DC, Fast-acting 500 mA 32VAC/50DC
Battery typeLithium ion
Battery capacityMinimum of five hours using new, fully charged battery with no alarms; optional 10-hour battery
Operating temperature5 °C to 40 °C (41 °F to 104 °F)
Transport/storage temperature-20 °C to 60 °C (-4 °F to 140 °F)
Operating humidity15% to 93% noncondensing
Operating altitude-170 m to 4877 m (-557 ft to 16,000 ft)
Weight1.5 kg (3 lbs)
Size82 H x 255 W x 155 D (mm), (3.23 H x 10.04 W x 6.10 D (in))
Available modesStandard, Homecare, and Sleep Study
ConnectivityWired and USB trend data export; nurse call capability

Frequently asked questions

What is the SpO2 measurement range of the PM100N?

The SpO2 measurement range is 1% to 100%.

What is the pulse rate measurement range?

The pulse rate measurement range is 20 to 250 beats per minute.

How long can the PM100N run on battery?

It runs for a minimum of five hours using a new, fully charged battery with no alarms; an optional 10-hour battery is available.

Which operating modes are available on the PM100N?

The available modes are Standard, Homecare, and Sleep Study.