This is a maintenance manual for the Nellcor N-200 pulse oximeter
This service manual covers test and repair of the Nellcor N-200 pulse oximeter, a portable instrument intended for continuously monitoring arterial oxygen saturation and pulse rate. The N-200 has a two-piece construction: a monitor that receives, processes and stores patient pulse data, and a powerbase that provides AC power to the monitor and drives peripheral display and recording devices. The two parts communicate over a bi-directional optical link.
The manual is addressed to qualified service personnel. It describes the instrument, its circuit-level theory of operation including the patient module, processor PCB, front panel display, power supply and battery charger PCBs and the powerbase, then gives troubleshooting and assembly procedures, monitor and powerbase disassembly, testing and calibration procedures with required test equipment, packing for shipment, spare parts lists and schematic diagrams. Warnings note that the device contains no user-serviceable parts and that servicing exposes hazardous voltages.
| Weight | about 8 pounds total |
|---|---|
| Battery operating time | up to 2 hours |
| Battery recharge time | fully recharged in 14 hours |
| Standard patient cable length | 4 meters |
| Maximum patient module leakage current | 10 microamperes |
| Trend memory | twelve hours of oxygen saturation and pulse rate data |
| Event memory | one hour of saturation, pulse rate, and pulse perfusion data |
| Number of PCBs in monitor | five printed circuit boards |
It is intended for continuously monitoring arterial oxygen saturation and pulse rate, measured noninvasively with a clip-on or adhesive-attached sensor on a finger or other site.
The front part is the monitor, which receives, processes and stores patient pulse data, and the back part is the powerbase, which provides AC power and drives peripheral display and recording devices.
Twelve hours of oxygen saturation and pulse rate data are stored in the trend memory, and one hour of data in the event memory.
No; the manual warns that the N-200 contains no user-serviceable parts and that servicing must be referred to qualified personnel.