Nellcor N395 SatSeconds Alarm Management Leaflet

The document describes the features of the Nellcor SatSeconds feature, which distinguishes clinically insignificant events from events of consequence, providing a smarter way to manage alarms.


Nellcor N395 SatSeconds Alarm Management Leaflet - cover page
Brand
Nellcor
Category
Health Care
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
187 KB
Published
03 August, 2012
Updated
28 September, 2023
Model
N-395
MD5 checksum
027677F8EE25529562482F84C9AFA77A

About this manual

The Nellcor N395 pulse oximeter includes a SatSeconds feature designed to manage nuisance alarms caused by minor and brief desaturation events. This feature distinguishes clinically insignificant events from those of consequence by calculating the product of the magnitude of the desaturation and the duration it persists below the alarm limit.

What's inside

  • SatSeconds Alarm Management
  • Nuisance Alarms
  • Desaturation Events
  • Alarm Thresholds
  • SatSeconds Limit Settings
  • Patient Safety

Frequently asked questions

What happens if a desaturation event is 2 points below the alarm limit for 3 seconds when the SatSeconds limit is set to 10?

No alarm is triggered because the event equals 6 SatSeconds, which is below the limit.

What is the minimum number of SatSeconds required to trigger an alarm when the limit is set to 10?

An alarm sounds in 2 seconds if the desaturation reaches 10 SatSeconds.