Molecular Devices ImageXpress Pico Automated Cell Imaging System Pre-Installation Guide

Pre-installation guide for the Molecular Devices ImageXpress Pico Automated Cell Imaging System covering environment, space, table, power, and computer…


Molecular Devices ImageXpress Pico Automated Cell Imaging System Pre-Installation Guide - cover page
Brand
Molecular Devices
Category
Other products
Document type
Installation Guide
Language
English
Pages
4
File format
PDF
File size
396 KB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
281E32B81220E6456247E869EB0EBD53

About this manual

This document provides the information needed to prepare for installation of the ImageXpress Pico Automated Cell Imaging System. It describes environmental, space, table, power, host computer, software, receiving, and installation requirements.

The system is designed for indoor laboratory use at 18°C to 30°C with 20% to 75% non-condensing humidity. Installation requires a sturdy table rated for at least 50 kg and an input voltage range of 100 VAC to 240 VAC.

What's inside

  • Environmental Requirements
  • Space and Table Requirements
  • Host Computer
  • Power Requirements
  • CellReporterXpress Software Requirements
  • Supported Browsers for Remote Clients
  • Facility Receiving Requirements
  • Installation Guidelines

Specifications

Operating temperature18°C to 30°C (64°F to 86°F)
Operating humidity20% to 75% non-condensing
Minimum distance from refrigerator5 m (16.4 ft)
Minimum distance from door2 m (6.5 ft)
Table weight ratingAt least 50 kg (110.2 lb)
Instrument weight without options38 kg (83.8 lb)
Instrument weight with Environmental Control System40 kg (88.2 lb)
Minimum table size94 cm (37.0 in.) wide by 61 cm (24.0 in.) deep
Rear clearance20 cm to 30 cm (7.9 in. to 11.8 in.)
Input voltage100 VAC to 240 VAC, 50/60 Hz, 4 amps maximum at 115 VAC
Voltage fluctuation toleranceWithin 10% of nominal voltage
Typical system power consumption452 watts
Host computer minimum RAM16 GB
Host computer display resolutionOptimized for 1920 x 1080 (Full HD)
Remote client minimum screen sizeAt least 9 inches
Ambient storage temperature-20°C to +60°C (-4°F to 140°F)
Ambient storage relative humidity15% to 75% (noncondensing)

Frequently asked questions

What environmental conditions are required for the ImageXpress Pico system?

It is designed to operate indoors under laboratory conditions at 18°C to 30°C with 20% to 75% non-condensing humidity.

What weight rating must the installation table have?

The system requires a sturdy table or lab bench with a weight rating of at least 50 kg (110.2 lb).

What are the minimum table dimensions?

The table or lab bench should be at least 94 cm (37.0 in.) wide by 61 cm (24.0 in.) deep.