Pre-installation guide for the Molecular Devices ImageXpress Pico Automated Cell Imaging System covering environment, space, table, power, and computer…
This document provides the information needed to prepare for installation of the ImageXpress Pico Automated Cell Imaging System. It describes environmental, space, table, power, host computer, software, receiving, and installation requirements.
The system is designed for indoor laboratory use at 18°C to 30°C with 20% to 75% non-condensing humidity. Installation requires a sturdy table rated for at least 50 kg and an input voltage range of 100 VAC to 240 VAC.
| Operating temperature | 18°C to 30°C (64°F to 86°F) |
|---|---|
| Operating humidity | 20% to 75% non-condensing |
| Minimum distance from refrigerator | 5 m (16.4 ft) |
| Minimum distance from door | 2 m (6.5 ft) |
| Table weight rating | At least 50 kg (110.2 lb) |
| Instrument weight without options | 38 kg (83.8 lb) |
| Instrument weight with Environmental Control System | 40 kg (88.2 lb) |
| Minimum table size | 94 cm (37.0 in.) wide by 61 cm (24.0 in.) deep |
| Rear clearance | 20 cm to 30 cm (7.9 in. to 11.8 in.) |
| Input voltage | 100 VAC to 240 VAC, 50/60 Hz, 4 amps maximum at 115 VAC |
| Voltage fluctuation tolerance | Within 10% of nominal voltage |
| Typical system power consumption | 452 watts |
| Host computer minimum RAM | 16 GB |
| Host computer display resolution | Optimized for 1920 x 1080 (Full HD) |
| Remote client minimum screen size | At least 9 inches |
| Ambient storage temperature | -20°C to +60°C (-4°F to 140°F) |
| Ambient storage relative humidity | 15% to 75% (noncondensing) |
It is designed to operate indoors under laboratory conditions at 18°C to 30°C with 20% to 75% non-condensing humidity.
The system requires a sturdy table or lab bench with a weight rating of at least 50 kg (110.2 lb).
The table or lab bench should be at least 94 cm (37.0 in.) wide by 61 cm (24.0 in.) deep.
The instrument requires an input voltage range from 100 VAC to 240 VAC, 50/60 Hz, with 4 amps maximum at 115 VAC.