Illumina MiSeq System Site Prep Guide provides specifications and guidelines for preparing a site for installation and operation of the MiSeq system.
This guide provides specifications and guidelines for preparing a site for installation and operation of the Illumina MiSeq system. It covers laboratory space requirements, electrical requirements, environmental constraints, computing requirements, and user-supplied consumables and equipment.
The MiSeq is shipped in one crate measuring 72.4 cm by 76.8 cm by 83.8 cm and weighing 90.7 kg. The installed instrument measures 68.6 cm by 52.3 cm by 56.5 cm, weighs 57.2 kg, and requires minimum clearances of 61 cm on each side, 10.2 cm at the rear, and 61 cm above.
| Crated width | 72.4 cm (28.5 in.) |
|---|---|
| Crated height | 76.8 cm (30.25 in.) |
| Crated depth | 83.8 cm (33 in.) |
| Crated weight | 90.7 kg (200 lbs.) |
| Instrument width | 68.6 cm (27 in.) |
| Instrument height | 52.3 cm (20.6 in.) |
| Instrument depth | 56.5 cm (22.2 in.) |
| Instrument weight | 57.2 kg (126 lbs.) |
| Side clearance | At least 61 cm (24 in.) |
| Rear clearance | At least 10.2 cm (4 in.) |
Only Illumina-authorized personnel can uncrate, install, or move the instrument.
Allow at least 61 cm (24 in.) on each side of the instrument.
A shielded CAT6 network cable is included.