Illumina TruSight Oncology 500 High Throughput Reference Guide

TruSight Oncology 500 High Throughput Reference Guide


Illumina TruSight Oncology 500 High Throughput Reference Guide - cover page
Brand
illumina
Category
Other
Document type
User Manual
Language
English
Pages
47
File format
PDF
File size
2.4 MB
Published
29 October, 2023
Updated
29 October, 2023
MD5 checksum
58D2DB08F1279CA4BF335965B7D2B265

About this manual

The TruSight Oncology 500 High Throughput (HT) protocol describes an enrichment-based approach to convert DNA and RNA extracted from formalin-fixed paraffin embedded (FFPE) tissue samples into libraries enriched for cancer-related genes that can be sequenced on Illumina sequencing systems. The kit is optimized to provide high sensitivity and specificity for low-frequency somatic variants across 523 genes, detecting single nucleotide variants, insertions, deletions, gene amplifications, multinucleotide variants, and immunotherapy biomarkers for tumor mutational burden and microsatellite instability in DNA. Fusions and splice variants are detected in RNA.

What's inside

  • Introduction
  • DNA/RNA Input Recommendations
  • Sample Qualification
  • DNA Shearing Recommendations
  • Library Prep DNA Only Workflow
  • Library Prep DNA and RNA Workflow
  • Denature and Anneal RNA
  • Synthesize First Strand cDNA
  • Synthesize Second Strand cDNA
  • Clean Up cDNA

Specifications

Minimum DNA input40 ng
Minimum RNA input80 ng
Optimal fragment size90-250 bp
Recommended sonication instrumentsCovaris E220evolution, LE220, or ME220
Recommended DNA QC methodIllumina FFPE QC Kit
Recommended RNA QC methodAdvanced Analytical Technologies Fragment Analyzer or Agilent Technologies 2100 Bioanalyzer
Recommended DNA QC thresholddelta Cq value < 5
Recommended RNA QC thresholdDV200 value of 20%

Frequently asked questions

What is the minimum DNA input required for the TruSight Oncology 500 HT assay?

The minimum DNA input required is 40 ng.

What is the minimum RNA input required for the TruSight Oncology 500 HT assay?

The minimum RNA input required is 80 ng.

What is the optimal fragment size for libraries prepared with the TruSight Oncology 500 HT assay?

The assay is optimized to prepare libraries from gDNA that are fragmented to 90-250 bp.