This user manual covers the operation, safety precautions, maintenance, and specifications of the HZJC-III fully automated insulating oil dielectric…
The document provides a comprehensive guide for the HZJC-III Insulating Oil Dielectric Strength Tester, manufactured by Wuhan Hanzhong Electric Equipment Manufacturing Co., Ltd. It details the instrument's features based on IEC 156 and GB 507-86 standards, including automatic testing, printing, and stirring functions controlled by a microcomputer.
The manual outlines specific technical specifications such as an AC220V power supply, a measurement range of 0-80KV, and a precision of 1%. It includes step-by-step instructions for panel operation, setting test parameters like the number of tests and timing, and interpreting results. Additionally, it covers essential safety precautions regarding high voltage and maintenance procedures to ensure proper instrument longevity. This manual is written in Chinese.
| Power Supply | AC220V ±10%, 50HZ ±5% |
|---|---|
| Measurement Range | AC 0-80KV |
| Limiting Current | 5mA |
| Voltage Rise Speed | 2KV/S ±10% |
| Stir Time Setting | 0-1 minute 39 seconds |
| Standstill Time Setting | 0-10 minutes 39 seconds |
| Operating Temperature | 0°C - 35°C |
| Relative Humidity | ≤75% |
| Oil Cup Gap | 2.5mm |
| Dimensions | 415×315×315mm³ |
| Weight | 28kg |
It uses a fully automatic magnetic vibrator to eliminate unevenness and air bubbles in the oil sample.
The instrument has a measurement range of AC 0 to 80KV.
The distance between the two electrodes must be adjusted to 2.5mm.
The device operates in temperatures from 0°C to 35°C with relative humidity at or below 75%.