The document provides a comprehensive guide for the HZJC-III Insulating Oil Dielectric Strength Tester, manufactured by Wuhan Hanzhong Electric Equipment Manufacturing Co., Ltd. It details the instrument's features based on IEC 156 and GB 507-86 standards, including automatic testing, printing, and stirring functions controlled by a microcomputer.

The manual outlines specific technical specifications such as an AC220V power supply, a measurement range of 0-80KV, and a precision of 1%. It includes step-by-step instructions for panel operation, setting test parameters like the number of tests and timing, and interpreting results. Additionally, it covers essential safety precautions regarding high voltage and maintenance procedures to ensure proper instrument longevity. This manual is written in Chinese.

What's inside

  • Introduction
  • Purpose and Features
  • Technical Specifications
  • Panel Description
  • Operating Steps
  • Safety Precautions
  • Maintenance and Care

Specifications

Power SupplyAC220V ±10%, 50HZ ±5%
Measurement RangeAC 0-80KV
Limiting Current5mA
Voltage Rise Speed2KV/S ±10%
Stir Time Setting0-1 minute 39 seconds
Standstill Time Setting0-10 minutes 39 seconds
Operating Temperature0°C - 35°C
Relative Humidity≤75%
Oil Cup Gap2.5mm
Dimensions415×315×315mm³
Weight28kg

Frequently asked questions

How does the instrument handle oil sample stirring?

It uses a fully automatic magnetic vibrator to eliminate unevenness and air bubbles in the oil sample.

What is the maximum voltage this tester can measure?

The instrument has a measurement range of AC 0 to 80KV.

What is the required gap between the electrodes in the oil cup?

The distance between the two electrodes must be adjusted to 2.5mm.

What are the operating temperature and humidity limits for this device?

The device operates in temperatures from 0°C to 35°C with relative humidity at or below 75%.