Instruction manual for Hanna Instruments HI9813-51 and HI9813-61 portable pH/EC/TDS/temperature meters, covering setup, calibration, measurement, and…
Hanna Instruments HI9813-51 and HI9813-61 are portable, water resistant meters designed for measuring pH, EC/TDS, and temperature when paired with their respective probes, HI1285-51 and HI1285-61. They use dedicated measurement-mode keys shown on the LCD and knobs for one-point calibration of pH and EC, with on-screen tutorial messages guiding calibration and measurement.
HI9813-61 adds CAL Check. The meters include a 3-in-1 combination probe with pH, EC/TDS, and temperature sensors and are described as best for hydroponics, greenhouses, farming, and ground water applications. The manual covers preliminary examination, specifications, battery replacement, sensor preparation, calibration, measurement, probe maintenance, accessories, certification, and warranty.
| pH range | 0.0 to 14.0 pH |
|---|---|
| pH accuracy | ±0.1 pH (@ 25 °C / 77 °F) |
| pH calibration | Manual, 1-point |
| EC range | 0.00 to 4.00 mS/cm |
| EC resolution | 0.01 mS/cm |
| EC accuracy | ±2% f.s. mS/cm (@ 25 °C / 77 °F) |
| EC calibration | Manual, 1-point |
| EC/TDS temperature compensation | Automatic from 0 to 70 °C with ß=2% / °C |
| TDS range | 0 to 1999 ppm |
| TDS accuracy | ±2% f.s. ppm (@ 25 °C / 77 °F) |
| TDS calibration | Manual, 1-point |
| TDS conversion factor | Variable, automatically adjusted from 0.56 to 0.78 based on actual EC readings |
| Temperature range | 0.0 to 70.0 °C |
| Included probe | HI1285-51 for HI9813-51; HI1285-61 for HI9813-61 |
| Battery life | Approximately 150 hours of continuous use |
| Auto shut-off | After 8 minutes of non-use |
| Battery type | 1 x 9V Alkaline |
| Operating environment | 0 to 50 °C (32 to 122 °F); 100% RH |
| Meter dimensions | 145 x 80 x 36 mm (5.7 x 3.1 x 1.4 in) |
| Meter weight | 230 g (8.1 oz.) |
| Probe reference | Single, Ag/AgCl |
| Probe junction | Cloth |
| Probe electrolyte | Gel |
| Probe maximum pressure | 0.1 bar |
| Probe range | pH: 0 to 13; EC: 0 to 6 mS/cm; Temperature: 0 to 50 °C (32 to 122 °F) |
| Probe tip shape | Spheric, Ø 8.0 mm (0.31 in) |
| Probe glass type | LT (low temperature) |
| Probe body material | Polypropylene |
| Probe cable | 1 m (3.3 ft) with 8-pin DIN connector |
| Probe dimensions | Length: 186 mm (7.32 in); shaft length: 95 mm (3.74 in); shaft diameter: 16 mm (0.62 in) |
| Probe plug | HI1285-51: DIN; HI1285-61: DIN with CAL Check |
Each meter is delivered in a rugged carrying case with its probe, pH 7.01 buffer, 1413 μS/cm conductivity solution, 1500 ppm TDS calibration solution, cleaning solution, beaker, 9V alkaline battery, quality certificate, and instruction manual; HI9813-61 also includes calibration check solution.
Frequent calibrations are recommended, and the pH range should be recalibrated whenever the electrode is replaced, at least once a month, after testing aggressive chemicals, or if HI9813-61 does not pass CAL Check.
Turn the instrument off, remove the three screws on the back to open the battery compartment, remove the old battery, insert one new 9V alkaline battery with correct polarity, and close the compartment with the three screws.
CAL Check is a feature of the HI9813-61 meter and its HI1285-61 probe; the manual lists CAL Check as HI9813-61 only.