The PACE 6000 is a modular pressure controller and indicator from GE, built on a modular platform and aimed at automated test systems and test-bench applications. The leaflet lists key features such as GE silicon piezoresistive and silicon resonance pressure sensing, pressure ranges up to 21 MPa gauge and absolute, 0.005% Rdg + 0.005% FS accuracy, 0.001% FS control stability, a 7-inch touchscreen with a graphical menu structure, and standard RS232 and IEEE connections.

Option modules described include programmable test programs, leak test, switch test (with electrical connections), analog output, relay contacts, burst test and an aviation version. The document then gives specifications for pressure measurement, display, performance, electrical data, environment and physical data, plus ordering information for the I6000 chassis, CM2 and CM2-B control modules, hardware accessories and mains leads. This manual is written in Chinese.

What's inside

  • Features
  • Programmable Test Programs
  • Leak Test
  • Switch Test
  • Analog Output
  • Burst Test
  • Relay Contacts
  • Aviation Version Option
  • Specifications
  • Ordering Information

Specifications

Standard pressure ranges35 kPa, 70 kPa, 100 kPa, 200 kPa, 350 kPa, 700 kPa, 1 MPa, 2 MPa, 3.5 MPa, 7 MPa, 10 MPa, 13.5 MPa, 17.2 MPa, 21 MPa
Gauge accuracy0.005% Rdg + 0.005% FS (includes nonlinearity, hysteresis, repeatability and temperature effects)
Control stability0.001% FS
Long-term stability200 kPa to 21 MPa: 0.01% reading/year
Display7 in (17.5 cm) TFT color VGA widescreen with integrated touchscreen
Analog output ranges0 to 10 V, 0 to 5 V, -5 to 5 V, 0/4 to 20 mA
CommunicationRS232, IEEE-488, SCPI, DPI515, DPI510 and DPI500
Power supply90 V AC to 130 V AC @ 47 to 63 Hz; 180 V AC to 260 V AC @ 47 to 63 Hz
Operating temperature10 °C to 50 °C (50 °F to 122 °F)
Pressure mediumDry, oil-free, non-corrosive gas; supply pressure should be 10% above maximum output pressure
WeightSingle control module 12.5 kg; dual control module 17.5 kg
Dimensions440 mm x 3U x 320 mm (17.3 in x 3U x 12.6 in)
EnclosureIP20

Frequently asked questions

What pressure ranges does the PACE 6000 cover?

Standard gauge ranges run from 35 kPa up to 21 MPa, and every gauge range includes negative pressure calibration as standard.

What is the gauge accuracy of the PACE 6000?

Gauge accuracy is 0.005% Rdg + 0.005% FS, including nonlinearity, hysteresis, repeatability and temperature effects.

Does the instrument use gas while it is measuring?

No. All supplied gas is used for system pressure control; there is no gas consumption in measurement mode or when switched off.

Which analog output signals can be programmed?

The output can be set to 0 to 10 V, 0 to 5 V, -5 to 5 V or 0/4 to 20 mA, proportional to the selected instrument range.