GE Druck PACE 6000 modular pressure controller and indicator: features, option modules, specifications and ordering information.
The PACE 6000 is a modular pressure controller and indicator from GE, built on a modular platform and aimed at automated test systems and test-bench applications. The leaflet lists key features such as GE silicon piezoresistive and silicon resonance pressure sensing, pressure ranges up to 21 MPa gauge and absolute, 0.005% Rdg + 0.005% FS accuracy, 0.001% FS control stability, a 7-inch touchscreen with a graphical menu structure, and standard RS232 and IEEE connections.
Option modules described include programmable test programs, leak test, switch test (with electrical connections), analog output, relay contacts, burst test and an aviation version. The document then gives specifications for pressure measurement, display, performance, electrical data, environment and physical data, plus ordering information for the I6000 chassis, CM2 and CM2-B control modules, hardware accessories and mains leads. This manual is written in Chinese.
| Standard pressure ranges | 35 kPa, 70 kPa, 100 kPa, 200 kPa, 350 kPa, 700 kPa, 1 MPa, 2 MPa, 3.5 MPa, 7 MPa, 10 MPa, 13.5 MPa, 17.2 MPa, 21 MPa |
|---|---|
| Gauge accuracy | 0.005% Rdg + 0.005% FS (includes nonlinearity, hysteresis, repeatability and temperature effects) |
| Control stability | 0.001% FS |
| Long-term stability | 200 kPa to 21 MPa: 0.01% reading/year |
| Display | 7 in (17.5 cm) TFT color VGA widescreen with integrated touchscreen |
| Analog output ranges | 0 to 10 V, 0 to 5 V, -5 to 5 V, 0/4 to 20 mA |
| Communication | RS232, IEEE-488, SCPI, DPI515, DPI510 and DPI500 |
| Power supply | 90 V AC to 130 V AC @ 47 to 63 Hz; 180 V AC to 260 V AC @ 47 to 63 Hz |
| Operating temperature | 10 °C to 50 °C (50 °F to 122 °F) |
| Pressure medium | Dry, oil-free, non-corrosive gas; supply pressure should be 10% above maximum output pressure |
| Weight | Single control module 12.5 kg; dual control module 17.5 kg |
| Dimensions | 440 mm x 3U x 320 mm (17.3 in x 3U x 12.6 in) |
| Enclosure | IP20 |
Standard gauge ranges run from 35 kPa up to 21 MPa, and every gauge range includes negative pressure calibration as standard.
Gauge accuracy is 0.005% Rdg + 0.005% FS, including nonlinearity, hysteresis, repeatability and temperature effects.
No. All supplied gas is used for system pressure control; there is no gas consumption in measurement mode or when switched off.
The output can be set to 0 to 10 V, 0 to 5 V, -5 to 5 V or 0/4 to 20 mA, proportional to the selected instrument range.