Routine maintenance guide for the Applied Biosystems PCR System 9600, covering well and heated cover cleaning, temperature verification, and diagnostic tests.
This routine maintenance document covers the Applied Biosystems GeneAmp PCR System 9600. It recommends performing maintenance at least 2-4 times per year, with a 3-month cycle for high-end users and a 6-month cycle for lower-end users to keep the instrument in top working order.
Procedures include cleaning the sample wells and heated cover with 100% isopropanol or a 10% bleach solution, rinsing with water, and running temperature verification and diagnostic tests. If a test fails, the instrument requires service by a qualified Applied Biosystems repair specialist.
| Recommended maintenance frequency | At least 2-4 times per year |
|---|---|
| Cool-down time before cleaning | 20-30 minutes |
| Sample well cleaning agent | 100% isopropanol |
It should be performed at least 2-4 times per year, with a 3-month cycle for high-end users and a 6-month cycle for lower-end users.
Wells are cleaned with a cotton swab soaked with 100% isopropanol, or a 10% bleach solution if contaminated, then rinsed with water.
A qualified Applied Biosystems repair specialist should service the instrument; in the U.S. call the Service Call Center, and in other countries contact the local Applied Biosystems service representative.