This document is an introduction to a product called 7300 Real-Time PCR System. It mainly introduces the features and functions of the product. The product can detect and quantify nucleic acid sequences without affecting data quality or dye selection flexibility.
The Applied Biosystems 7300 Real-Time PCR System is an affordable platform for detecting and quantifying nucleic acid sequences using thermal cycling, fluorescence detection, and application-specific software. It supports multiple applications including gene expression analysis, SNP genotyping, and pathogen quantitation with four-color detection flexibility. The system runs in under two hours using 96-well plates or tubes with a 20–100 μL volume range. It includes precision optics, a CCD camera, and powerful software for data analysis and visualization. The system can be used as a plate reader or thermal cycler and comes with a Dell computer for operation.
| Thermal Cycling System | Peltier-based, 96-well block |
|---|---|
| Temperature Range | 4°C – 100°C |
| Temperature Accuracy | ±0.25°C of setpoint/display temperature measured at 3 minutes after clock start |
| Temperature Uniformity | ±0.50°C, 30 seconds after clock start |
| Supported Volumes | 20 – 100 μL |
| Quantitative PCR Run Time | < 1 hour 50 minutes |
| Fluorescence Detection | Four-color detection with FAM, VIC, NED, TAMRA, ROX dyes |
| Optical System | Single excitation, four emission filters, and CCD camera |
| Data Collection | Data collected in all 4 filters for all wells regardless of plate setup |
| Dynamic Range | 9 logs |
It supports gene expression analysis, SNP genotyping, pathogen quantitation, isothermal assays, and plus/minus assays with internal positive controls.
Quantitative PCR runs are completed in under two hours, with a run time of less than 1 hour 50 minutes.
The system includes powerful, user-friendly software with plate setup wizards, real-time monitoring, auto-baseline and auto-threshold, and customizable graphing options.
The temperature accuracy is ±0.25°C of setpoint/display temperature measured at 3 minutes after clock start.