This document is the installation manual of the Applied Biosystems® 3500/3500xL Genetic Analyzer, which introduces the preparation work required for the installation of the device.
This guide provides step-by-step instructions for preparing your site to install the Applied Biosystems 3500/3500xL Genetic Analyzer. It includes a site preparation checklist, installation workflow, and training information. The document outlines customer responsibilities, space and electrical requirements, and safety precautions. Installation typically takes 1.5 business days and includes basic operator training by a Life Technologies service representative.
Key requirements include adequate bench space, environmental controls (15–30°C, 20–80% humidity), and proper electrical outlets (AC 100–240 V, 50/60 Hz). The instrument must be installed in a BSL-3 or BSL-4 free area, and power must remain accessible for safety. A power line regulator is recommended in areas with voltage fluctuations beyond ±10%.
| Instrument Power | AC 100-240 V ±10%, 50/60 Hz, 3.1 A, 320 VA |
|---|---|
| Computer Power | AC 100-240 V ±10%, 50/60 Hz, 2.1 A, 125 VA |
| Monitor Power | AC 100-240 V ±10%, 50/60 Hz, 1.5 A, 65 VA |
| Operating Temperature | 15 to 30°C (59 to 86°F) |
| Operating Humidity | 20 to 80% relative humidity, noncondensing |
| Instrument Dimensions (Closed Door) | 72 cm (28.3 in.) height, 61 cm (24 in.) depth, 61 cm (24 in.) width |
Installation and testing takes about 1.5 business days.
A 1.5-kVA power line regulator is recommended for areas with voltage fluctuations beyond ±10% of nominal voltage.