Quick guide for using Agilent's PlateLoc thermal microplate sealer to seal microplates effectively with tips on materials, parameters, and seal evaluation.
This quick reference guide helps users operate the Agilent PlateLoc Thermal Microplate Sealer to seal microplates effectively. It covers selecting compatible labware, optimizing sealing parameters like temperature and time, and evaluating seal quality. The guide advises using polypropylene microplates and avoiding acrylic ones. It also explains how to inspect seals using visual indicators on the seal material.
The guide is intended for use alongside the full user manual and seal selection guide. It includes safety notices, warranty disclaimers, and contact information for technical support and customer service.
Use polypropylene microplates because they are the easiest to seal effectively and reliably. Avoid acrylic microplates because they are not compatible with thermal microplate sealing.
After sealing, inspect the underside of the removed seal material. If there are unbroken impressions of each well, the microplate was properly sealed.