The Agilent Intuvo 9000 GC System represents a new breed of gas chromatography designed to offer advanced technologies such as direct heating, planar column design, and fast column changes. The system features a user-friendly touchscreen interface and supports a wide range of inlets and detectors, including FID, TCD, ECD, NPD, FPD, SCD/NCD, and mass spectrometers.

What's inside

  • Chromatographic performance
  • System capabilities
  • Column oven
  • Electronic pneumatics control
  • Inlets
  • Detectors
  • Flame ionization detector
  • Thermal conductivity detector
  • Mass spectrometers

Specifications

Retention time repeatability<0.008% or <0.0008 minutes
Area repeatability<1% RSD
Oven operating temperature range+10 °C to 450 °C
Maximum achievable temperature ramp rate250 °C/min
Pressure setpoint control precision0.001 psi
FID minimum detectable level<1.4 pg C/s
TCD minimum detectable level400 pg tridecane/mL
Micro-ECD minimum detectable level<4.4 fg/mL lindane

Frequently asked questions

What detectors does the Agilent Intuvo 9000 support?

The system supports FID, TCD, ECD, NPD, FPD, SCD/NCD, and Mass Spec (single and triple quad).

What is the maximum oven temperature?

The oven accommodates a maximum operating temperature of 450 °C.

Does the system include automatic leak checking?

Yes, automatic leak checking can be enabled by the user for every run or during maintenance.