Agilent user guide for InfinityLab Poroshell 120, ZORBAX, Pursuit and Polaris reversed-phase HPLC columns, covering installation, conditioning, operating…
This Agilent column user guide provides general information for InfinityLab Poroshell 120, ZORBAX, Pursuit and Polaris reversed-phase columns. It covers installation, column conditioning, operating tips, mobile phase selection, column care, cleaning, storage and ways to obtain better chromatographic results.
The guide includes recommended pH ranges and maximum operating temperatures by phase, maximum operating pressures by column type and particle size, recommended starting gradients, and notes on guard columns, fittings and safety considerations. This manual is written in Chinese.
| InfinityLab Poroshell 120 and Poroshell HPH, 1.9 µm particle size pressure limit | 1300 bar (19500 psi) |
|---|---|
| InfinityLab Poroshell 120 and Poroshell HPH, 2.7 µm and 4 µm particle size pressure limit | 600 bar (9000 psi) |
| ZORBAX RRHD, 1.8 µm particle size pressure limit | 1300 bar (19500 psi) |
| ZORBAX RRHT, 1.8 µm particle size pressure limit | 600 bar (9000 psi) |
| Other ZORBAX, Rx, Pursuit, Polaris, HC, TC and other bonded phases pressure limit | 400 bar (6000 psi) |
| Recommended pH range for EC-C18, EC-C8, Bonus-RP, Eclipse Plus, XDB and HC-C18(2)/TC-C18(2) phases | pH 2.0 to 9.0 |
| Maximum operating temperature for EC-C18, EC-C8, Bonus-RP, Eclipse Plus, XDB and HC-C18(2)/TC-C18(2) phases | 60 °C |
| Recommended pH range for Poroshell 120 PFP, Phenyl-Hexyl, XDB-CN, Eclipse PAH, ODS, Phenyl and other bonded phases | pH 2.0 to 8.0 |
| Recommended pH range for StableBond SB phases and ZORBAX Rx-C8 | pH 1.0 to 8.0 |
| Maximum operating temperature for StableBond C18 and Poroshell 120 SB-C18 | 90 °C |
| Maximum operating temperature for Rx-C8, SB-Aq, SB-C3 and SB-Phenyl | 80 °C |
| Recommended pH range for InfinityLab Poroshell HPH-C18 and HPH-C8 | pH 3.0 to 11.0 |
| Recommended pH range for ZORBAX Extend C18 | pH 2.0 to 11.5 |
| Inlet frit nominal porosity for columns with particle size of 2.7 µm or greater | 2 µm |
| Recommended tubing internal diameter for Fast LC/high efficiency columns | 0.12 mm id red or 0.075 mm id black |
It provides general information for all InfinityLab Poroshell 120, ZORBAX, Pursuit and Polaris columns.
No. Every column is tested before shipment and shipped in the test eluent, so rinsing with water is not necessary for first use.
ZORBAX RRHT and ZORBAX RRHD 1.8 µm columns can only be operated in the direction of flow marked on the column.
Silica-based bonded phase columns should be stored in a pure organic solvent, with any buffer first removed by purging, and the end-fittings tightly capped with end-plugs.
No. For columns with particles smaller than 1.8 µm, do not backflush the column; replace it instead.