This manual provides instructions for installing, configuring, and setting up parameters for the Agilent InfinityLab LC with ISET system to enable…
The Agilent InfinityLab LC with ISET (Intelligent System Emulation Technology) allows for the transfer of HPLC and UHPLC methods between different laboratory instruments without changes in retention time and resolution. This technology addresses design differences between LC systems, such as delay volume and gradient mixing behavior, which typically affect method transferability.
| Retention Time (RT) Deviation | ±5% or ±0.3 min (for RT > 6 min) |
|---|---|
| Retention Time Deviation | ±5% or ±0.1 min (for ∆RT ≤ 2 min) |
ISET is used for instrument-to-instrument method transfer, allowing legacy methods to be performed on modern instruments without changes in retention time and resolution.
Design differences such as power range, delay volume, mixing behavior, temperature control, extra column volume, and detector cell design affect method transfer.