Maintenance and repair guide for Agilent InfinityLab LC Series Multisamplers, covering component replacement and safety procedures.
This service manual provides step-by-step instructions for maintaining and repairing the Agilent InfinityLab LC Series Multisamplers. It includes procedures for replacing parts such as needle assemblies, rotor seals, and peristaltic pump cartridges. Safety warnings are included for electrical hazards, sharp edges, and hazardous materials. The manual also covers cleaning, front door installation, and cable identification for connectivity.
| Change needle/needle seat | 60000 needle into seat |
|---|---|
| Change peristaltic pump cartridge | 3000 min on time |
| Change rotor seal | 30000 injections |
Do not remove the cover of the module. Only certified persons are authorized to carry out repairs inside the module. Be careful when getting in contact with sharp metal areas.
The needle/needle seat should be changed every 60,000 injections.
The peristaltic pump cartridge should be replaced every 3000 minutes of on-time use.