User's guide for the Agilent E3632A DC power supply covering safety, operation, remote sensing, programming, and specifications.
The Agilent E3632A User's Guide describes the Agilent Technologies E3632A DC power supply. It includes safety notices, operating instructions, remote voltage sensing, load considerations, remote programming, and characteristics and specifications.
The guide also covers cooling and bench operation, output rating and noise, stability with remote sensing, series and parallel connections, and environmental and electrical specifications.
| Operating temperature | 0 °C to 40 °C (full rated output) |
|---|---|
| Storage temperature | –20 °C to 70 °C |
| Humidity | Up to 80% RH |
| Altitude | Up to 2000 m |
| Safety class | Safety Class I |
| Installation category | II (for indoor use) |
| Pollution degree | 2 |
It operates at full rated output from 0 °C to 40 °C.
No cleaning is required; use a dry cloth to remove dust from the enclosure.
Twist the sense leads and run them parallel and close to the load leads; shield them in noisy environments.
The Unreg annunciator turns on to indicate that the output is unregulated when excessive demand causes the supply to lose regulation.