A checklist and information guide for preparing the laboratory site, including space, environmental conditions, power, and gas supply requirements for the…
This document serves as a site preparation checklist and information guide for the Agilent 7820A Gas Chromatograph (GC). It outlines the necessary laboratory space, environmental conditions, power requirements, and gas supply specifications to ensure a successful and timely installation. The guide emphasizes that correct site preparation is the key first step in ensuring the instrument and software systems operate reliably over an extended lifetime.
| Operating Temperature Range (Recommended) | 15 to 35 °C |
|---|---|
| Carrier Gas Purity | 99.9995% or better |
The recommended operating temperature range is 15 to 35 °C.
A simple system that includes a GC and a computer requires at least 106 cm of bench space.
No, power line conditioners should not be used with the Agilent 7820A GC.
Agilent recommends a carrier and detector gas purity of 99.9995% or better.