This guide details site preparation requirements for the Agilent 7800/7900 ICP-MS, covering space, environmental conditions, power, cooling water, gas, and…
This document serves as a checklist and information guide to ensure successful installation of the Agilent 7800/7900 Inductively Coupled Plasma Mass Spectrometer (ICP-MS). It outlines critical pre-installation specifications including laboratory space dimensions, environmental conditions such as temperature and humidity, and utility requirements. This manual is written in Chinese.
| Operating Temperature Range | 15-30 ºC (59-86 ºF) |
|---|---|
| Temperature Variation Limit | < 2 ºC/h |
| Relative Humidity Range | 20-80 % |
| Maximum Altitude | 2000 meters |
| Required Clearance Around Instrument | 60 cm or 2 feet |
The customer is responsible for ensuring the site meets pre-installation specifications, including laboratory space, environmental conditions, power requirements, cooling water, gas, and network configuration before the installation date.
Yes, a suitable exhaust system that functions correctly is mandatory; otherwise, Agilent service providers will not install the system.
The document specifies that cooling water is required, with detailed running conditions and heat removal capabilities (up to 4,500 W via cooling water and exhaust) described in specific sections.