The Agilent 7010B Triple Quadrupole GC/MS System data sheet lists specifications for the Agilent 7890B GC and 7010B MS, including IDL and MRM performance.
This data sheet describes the Agilent 7010B Triple Quadrupole GC/MS System, which combines the Agilent 7890B Gas Chromatograph with the Agilent 7010B Mass Spectrometer for applications such as pesticides and environmental pollutants.
It lists mass spectrometer parameters including EI standard operation with optional CI, a mass range of m/z 10 to 1,050, and MRM speed up to 800 transitions per second. The document also provides GC specifications, integrated system features, instrument dimensions, and installation checkout specifications such as the 0.5 fg octafluoronaphthalene IDL.
| Mode of operation | EI standard, CI optional |
|---|---|
| Ion source temperature | 150 to 350 °C |
| Mass range | m/z 10 to 1,050 |
| MRM speed | 800 transitions/sec |
| Minimum MRM dwell | 0.5 msec |
| Collision energy | Selectable up to 60 eV |
| Oven temperature | Ambient + 4 to 450 °C |
| Instrument Detection Limit | 0.5 fg or less octafluoronaphthalene (OFN) |
The EI MRM IDL is 0.5 fg or less octafluoronaphthalene (OFN).
EI is standard and CI is optional.
It uses Agilent MassHunter acquisition, data handling, and reporting software.
The mass range is m/z 10 to 1,050.