Agilent AriaMx SYBR Qualification Plate Kit 5190-7708 Protocol

Protocol for running a SYBR qualification test on the AriaMx Real-Time PCR System using the AriaMx SYBR Qualification Plate Kit.


Agilent AriaMx SYBR Qualification Plate Kit 5190-7708 Protocol - cover page
Brand
Agilent
Category
Other
Document type
User Manual
Language
English
Pages
13
File format
PDF
File size
174 KB
Published
16 November, 2026
Updated
16 November, 2026
MD5 checksum
9B295D1F5858C5ADFD7948679CF5C21B

About this manual

This protocol provides instructions for using the AriaMx SYBR Qualification Plate Kit to run a SYBR qualification test on the AriaMx Real-Time PCR System. The kit contains one 96-well plate whose wells are pre-filled with a QPCR master mix that includes the SYBR Green dye.

The guide covers instrument optical module requirements, plate storage and handling, plate preparation, running the test in the AriaMx software, and reviewing the qualification test graphical data.

What's inside

  • Before You Begin
  • Overview
  • Instrument Optical Module Requirements
  • Kit components
  • Plate storage and handling
  • Protocol
  • Prepare the plate
  • Run the test
  • About the Qualification Test graphical data
  • What's New in Version D1

Specifications

Part Number5190-7708
Kit contentsOne 96-well plate
Reagent volume per well20 µL of QPCR reagents
Storage temperature–20°C upon receipt
Thawing time at room temperature5–15 minutes
Plate centrifugation500 × g for at least one minute
Required optical modulesSYBR/FAM optical module and a ROX optical module
Pass criterionBoth populations at least 98.5% distinct after outlier wells are culled

Frequently asked questions

What instrument and optical modules are required to run the qualification test?

The test is run on the AriaMx Real-Time PCR System, which must be equipped with a SYBR/FAM optical module and a ROX optical module.

How should the AriaMx SYBR Qualification Plate be stored and handled?

The plate is shipped on dry ice and should be stored at –20°C upon receipt. Wear chemical-resistant powder-free gloves whenever handling it.

How is the plate prepared before running the qualification test?

Thaw the plate at room temperature for 5–15 minutes without removing the seal, invert and re-invert it five times to mix the wells, then centrifuge at 500 × g for at least one minute and check for bubbles.

How do I know whether the qualification test passed?

In the AriaMx software, open the Graphical Displays panel on the right side of the screen. A passed test displays Results Pass; a failed test displays Results Check.