Protocol for running a SYBR qualification test on the AriaMx Real-Time PCR System using the AriaMx SYBR Qualification Plate Kit.
This protocol provides instructions for using the AriaMx SYBR Qualification Plate Kit to run a SYBR qualification test on the AriaMx Real-Time PCR System. The kit contains one 96-well plate whose wells are pre-filled with a QPCR master mix that includes the SYBR Green dye.
The guide covers instrument optical module requirements, plate storage and handling, plate preparation, running the test in the AriaMx software, and reviewing the qualification test graphical data.
| Part Number | 5190-7708 |
|---|---|
| Kit contents | One 96-well plate |
| Reagent volume per well | 20 µL of QPCR reagents |
| Storage temperature | –20°C upon receipt |
| Thawing time at room temperature | 5–15 minutes |
| Plate centrifugation | 500 × g for at least one minute |
| Required optical modules | SYBR/FAM optical module and a ROX optical module |
| Pass criterion | Both populations at least 98.5% distinct after outlier wells are culled |
The test is run on the AriaMx Real-Time PCR System, which must be equipped with a SYBR/FAM optical module and a ROX optical module.
The plate is shipped on dry ice and should be stored at –20°C upon receipt. Wear chemical-resistant powder-free gloves whenever handling it.
Thaw the plate at room temperature for 5–15 minutes without removing the seal, invert and re-invert it five times to mix the wells, then centrifuge at 500 × g for at least one minute and check for bubbles.
In the AriaMx software, open the Graphical Displays panel on the right side of the screen. A passed test displays Results Pass; a failed test displays Results Check.