This guide details the site preparation requirements, safety practices, and facility specifications for installing the Agilent 4100 Microwave Plasma Atomic…
This document provides the necessary information to successfully prepare a site for the installation of an Agilent 4100 Microwave Plasma Atomic Emission Spectrometer (MP-AES). It covers safety practices, laboratory environment requirements, instrument shipping information, and laboratory facilities including electrical and gas needs. Upon completion of the site preparation, a checklist should be filled out and sent to Agilent to arrange installation.
| Frequency | 50–60 Hz ±1 Hz |
|---|---|
| Power rating | 2.04 kVA |
| Argon purity | 99.0 % |
| Permissible pressure range (Argon) | 200–440 kPa (29–63 psi) |
| Required flow rate (Plasma Gas) | 25 SLPM |
| Connection hardware | 1/4 in Swagelok |
Allow a minimum of four hours for installation by an Agilent field service engineer, or up to eight hours if accessories are included.
A personal computer with Windows 7 64-bit operating system installed is required.
The recommended regulated supply pressure for Argon is 210 kPa (30 psi).