This technical note provides instructions for installing, connecting, and using the Agilent 1260 Infinity II Manual Preparative Injector.
This document serves as a technical guide for the Agilent 1260 Infinity II Manual Preparative Injector. It details the unpacking process, installation steps including the mounting of the holder and ring stand, and the configuration of flow connections and internal reducers. The guide also covers safety warnings regarding solvent handling and leak drainage.
| Injection seal material | PEEK |
|---|---|
| pH compatibility | pH 0 – 14 |
| Sample loop volume limit for ejection warning | larger than 100 µL |
Use at least two to three times the loop volume, for example 40 – 60 μL of sample for a 20 μL sample loop.
Leave the syringe in the needle port and turn the handle to the INJECT position before removing it.