Thermo D10P-06 Low Sodium Measurement User Manual

This document introduces the low sodium measurement method for D10P-06 kit. The kit includes Dual Star dual channel pH/ISE benchtop meter, 8411BN sodium ion electrode, 800500U Ross reference electrode and other equipment and solutions. The preparation method of calibration standard solution and the storage requirements of electrodes are provided. The measurement precautions include using high-quality plastic containers, using fresh standard solutions, avoiding interference from hydrogen ions, etc. Suggestions for using direct measurement method and known addition method are given, and the importance of reducing sample-air contact is emphasized.


Thermo D10P-06 Low Sodium Measurement User Manual - cover page
Brand
Thermo
Category
Other
Document type
User Manual
Language
Chinese
Pages
4
File format
PDF
File size
127 KB
Updated
29 September, 2023
MD5 checksum
792D38F7598167750F3E97AE1800001E

About this manual

The D10P-06 kit provides a method for low‑sodium analysis in power‑plant boiler water and steam. It uses a separate half‑cell sodium ion electrode (8411BN) together with a Ross reference electrode (800500U) and includes all necessary accessories such as temperature‑compensation electrodes and storage solutions.

The manual details the preparation of sodium standard solutions (PN4, PN5, PN6), storage and cleaning of electrodes, instrument setup on a Dual Star pHISE meter, calibration steps, and the measurement protocol. It also lists safety and handling precautions to ensure accurate and repeatable results. This manual is written in Chinese.

What's inside

  • Introduction
  • Required Equipment and Solutions
  • Standard Solution Preparation
  • Electrode Storage
  • Measurement Precautions
  • Instrument Setup
  • Calibration and Measurement Procedure
  • Operation Manual

Specifications

Sodium ion electrode model8411BN
Reference electrode model800500U
Temperature compensation electrode model927007MD
Electrode storage solution dilution ratio1:100
Measurement resolution3 significant figures

Frequently asked questions

How often should the sodium ion electrode storage solution be replaced?

The storage solution should be changed weekly, or sooner if crystals appear or the solution becomes contaminated.

What pH is required to avoid interference from oxygen ions during measurement?

The sample pH should be greater than 10 to prevent oxygen ion interference.

What is the recommended dilution ratio for the sodium electrode storage solution?

Mix 1 part of the storage solution (841101) with 100 parts of de‑ionized water (1:100).

How many significant figures does the instrument display for sodium concentration?

The instrument displays results with three significant figures.