This manual introduces the types of conductivity electrodes, their usage and calibration methods.
Thermo DureProbe 2/4 conductivity electrode manual describes 2-ring and 4-ring conductivity cells for laboratory and field use, including glass, stainless steel, and epoxy body variants with their cell constants and recommended measurement ranges. It also covers selecting an electrode for the sample conductivity range.
The manual explains automatic and manual calibration with conductivity standard solutions, lists required equipment and accessories, and gives cleaning, storage, and troubleshooting guidance for DureProbe electrodes. This manual is written in Chinese.
| Cell constant, 9902BN glass/platinum 2-ring electrode | 0.1 cm^-1 |
|---|---|
| Recommended measurement range, 9902BN | 0.1 μS/cm to 100 mS/cm |
| Cell constant, 013016MD stainless steel 2-ring electrode | 0.1 cm^-1 |
| Recommended measurement range, 013016MD | 0.01 μS/cm to 300 mS/cm |
| Cell constant, 011050MD epoxy/platinum 2-ring electrode | 1 cm^-1 |
| Recommended measurement range, 011050MD | 1 μS/cm to 20 mS/cm |
| Cell constant, 013005MD/013010MD/013025MD epoxy/graphite 4-ring electrodes | 0.475 cm^-1 |
| Recommended measurement range, 013005MD/013010MD/013025MD | 1 μS/cm to 200 mS/cm |
| Cell constant, 013150A/013300A epoxy/graphite 4-ring electrodes with protection sleeve | 0.475 cm^-1 |
| Recommended measurement range, 013150A/013300A | 1 μS/cm to 200 mS/cm |
| Cell constant, 013610MD epoxy 4-ring electrode | 0.55 cm^-1 |
| Recommended measurement range, 013610MD | 10 μS/cm to 200 mS/cm |
| Cell constant, 011510MD epoxy/graphite 2-ring electrode | 1 cm^-1 |
| Recommended measurement range, 011510MD | 10 μS/cm to 200 mS/cm |
| Cell constant, 018020MD glass/platinum 2-ring electrode | 10 cm^-1 |
| Recommended measurement range, 018020MD | 10 μS/cm to 2000 mS/cm |