Thermo Scientific NanoDrop 2000/2000C Spectrophotometer User Manual

The NanoDrop 2000/2000C spectrophotometer is a spectrophotometer produced by Thermo Scientific. It can detect 0.5-2ul samples with high accuracy and repeatability. ND2000C not only provides the convenience of NanoDrop sample retention patented technology, but also can use traditional cuvettes to detect samples.


Thermo Scientific NanoDrop 2000/2000C Spectrophotometer User Manual - cover page
Brand
Thermo
Document type
User Manual
Language
Chinese
Pages
54
File format
PDF
File size
2.3 MB
Published
17 December, 2011
Updated
30 September, 2023
MD5 checksum
4AC1C10B571BCB33E764D1E1D8B916F0

About this manual

The NanoDrop 2000/2000C spectrophotometer from Thermo Scientific provides micro‑volume UV‑Vis measurements with a wavelength range of 190‑840 nm. It supports both the proprietary sample‑holding pedestal and traditional cuvette detection, allowing analysis of nucleic acids, proteins and other biomolecules with sample volumes as low as 0.5 µL.

The manual explains hardware installation, USB connection, software installation on Windows XP/Vista, sample loading procedures, blank calibration, and specific application modes such as protein Pierce 660 nm and cell culture measurements. Maintenance tips, cleaning instructions and troubleshooting guidance are also included. This manual is written in Chinese.

What's inside

  • Instrument Introduction
  • Specifications
  • Sample Pedestal Loading
  • Cuvette Detection
  • Software Installation
  • USB Connection
  • Blank Calibration
  • Fluorescent Dye Applications
  • Cell Culture Measurements
  • Maintenance and Cleaning

Specifications

Instrument typeSpectrophotometer
Minimum sample volume0.5 µL
Wavelength range190‑840 nm
Light sourceXenon flash lamp
Detector2048‑pixel linear silicon CCD array
Wavelength accuracy±1 nm
Spectral resolution≤1.8 nm (FWHM @ 253.7 nm Hg)
Detection limit (dsDNA)2 ng/µL
Maximum concentration (dsDNA)15 000 ng/µL
Detection time< 5 s (pedestal), < 3 s (cuvette)
Power consumption12‑18 W (max 30 W)

Frequently asked questions

What is the minimum sample volume required for the NanoDrop 2000/2000C?

The pedestal mode can measure as little as 0.5 µL, while the cuvette mode typically uses 1‑2 µL of sample.

What wavelength range does the NanoDrop 2000/2000C cover?

It operates over a full UV‑Vis range from 190 nm to 840 nm.

How do I perform a blank calibration in pedestal mode?

Load a blank onto the pedestal, lower the arm, click the Blank button, and repeat until the absorbance variation is less than 0.04 A.

Which operating systems are supported by the NanoDrop software?

The software runs on 32‑bit Windows XP and Windows Vista.