Zhiyuan Electronics iCAN-6605 6-Channel Thermocouple Measuring Module User Manual

iCAN-6605 6 Channel Thermocouple Temperature Module


Brand
ZLG
Document type
User Manual
Language
Chinese
Pages
41
File format
PDF
File size
2.1 MB
Updated
30 September, 2023
MD5 checksum
924565A2E2B22E16FA14141D8078F79E

About this manual

The iCAN-6605 is a 6‑channel thermocouple measuring module designed for industrial temperature data acquisition. It supports eight standard thermocouple types, provides 0.1 °C resolution, and includes five digital output channels for status indication or user control.

The manual covers functional overview, technical specifications, interface description, power requirements, CAN baud‑rate and MAC ID configuration (via DIP switch or network), LED indicators, wiring diagrams, communication message formats, application examples, and safety precautions. This manual is written in Chinese.

What's inside

  • Function Overview
  • Technical Specifications
  • Interface Description
  • Power Supply
  • CAN Baud Rate and MAC ID Settings
  • Signal Indicator LEDs
  • CAN Bus Connection
  • Power and Communication Wiring
  • Thermocouple Input
  • Analog Input

Specifications

Input channels6 differential, configurable for thermocouple, V, mV, mA
Digital output channels5 open‑drain channels
Digital output max load30 V, 30 mA
Supported thermocouple types and rangesJ –210 °C~1200 °C, K –200 °C~1370 °C, E –100 °C~1000 °C, T –200 °C~400 °C, N –200 °C~1300 °C, B 650 °C~1800 °C, R 0 °C~1750 °C, S 0 °C~1760 °C
Analog input range±2.5 V or ±20 mA (with 125 Ω resistor)
Temperature resolution0.1 °C
Bandwidth5.24 Hz
Sampling rate1 sample / s (6 channels per sample)
Cycle transmission interval10 ms – 2.55 s
Isolation voltage1000 V DC
Power consumption1500 mW @ 12 V DC
Operating temperature range-25 °C – 70 °C

Frequently asked questions

What is the maximum load the digital output channels can drive?

Each digital output channel can source up to +30 V with a maximum current of 30 mA.

Which thermocouple types are supported by the module?

The iCAN-6605 supports J, K, E, T, N, B, R, and S thermocouples with the corresponding temperature ranges listed in the manual.