This Micro Motion service bulletin (MMI-SB-128, Rev A, February 2016) describes the pressure relief feature added to model CNG050 Coriolis flowmeters in 2006. The feature evacuates the sensor case if primary containment is lost.
The CNG050 case is not pressure-rated and has a typical burst pressure of 1,530 psi (105 bar). Sensors with a tag stamped “2006” or higher include the pressure relief feature, located underneath the calibration tag.
| Typical burst pressure | 1,530 psi (105 bar) |
|---|---|
| Pressure relief feature location | Underneath the calibration tag |
| Year pressure relief feature added | 2006 |
| Tag year indicating pressure relief feature | 2006 or higher |
If the sensor tag is stamped with “2006” or higher, the CNG050 includes the pressure relief feature.
The typical burst pressure is 1,530 psi (105 bar).
Micro Motion added the pressure relief feature to the CNG050 in 2006.
It is located underneath the calibration tag.