This product is a device for measuring the concentration of various ions in water, with 8 types of selection, including ammonium ions, nitrite, nitrate, magnesium ions, sodium ions, potassium ions, chloride ions, and calcium ions.
This manual describes an all-in-one sensor for measuring ion concentrations in water, available in eight ion types: ammonium, nitrite, nitrate, magnesium, sodium, potassium, chloride, and calcium. It is used for online automatic continuous analysis in industrial wastewater, surface water, drinking water, seawater, and industrial process control.
The document covers hardware connection, installation, configuration software use, ModBus-RTU communication protocol and registers, electrode calibration, and maintenance precautions. It also lists technical parameters such as power supply, measurement ranges, accuracy, response time, operating conditions, cable length, and electrode service life. This manual is written in Chinese.
| Power supply | DC 10~30V |
|---|---|
| Power consumption | 0.19W |
| Communication interface | RS485, ModBus-RTU |
| Default baud rate | 4800 bit/s |
| Supported baud rates | 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 bit/s |
| Ion concentration range (100 mg/L type) | 0-100.00 mg/L, resolution 0.01 mg/L (chloride not supported) |
| Ion concentration range (1000 mg/L type) | 0-1000.0 mg/L, resolution 0.1 mg/L (sodium, magnesium, calcium only) |
| Ion concentration range (3500 mg/L type) | 0-3500.0 mg/L, resolution 0.1 mg/L (chloride only) |
| Ion measurement error | 5% FS |
| Repeatability error | ±1 mg/L |
| Response time | <60 s |
| Operating temperature | 0-50 °C |
| Pressure rating | 0.2 MPa |
| Cable length | Default 5 m; 10 m, 15 m, and 20 m customizable |
| Electrode service life | 3-6 months |
| Modbus address range | 1-254, default 1 |
It is available in eight ion types: ammonium, nitrite, nitrate, magnesium, sodium, potassium, chloride, and calcium.
The default baud rate is 4800 bit/s.
Theoretically, up to 254 485 sensors can be used on one bus.
Unused electrodes should be stored in diluted standard solutions of the respective ions.