This document is a basic 1D NMR spectroscopy training guide for Varian 300 MHz NMR spectrometers.
This guide covers basic spectrometer operation for Varian 300 MHz NMR spectrometers, including the Inova-300-1 with a 5mm 4-Nucleus Plus probe and the Inova-300-2 with a 5mm Autoswitchable probe. It provides step-by-step instructions for logging in, changing samples, acquiring and processing 1H, 13C, 19F and 31P spectra, and logging out.
It also includes supplementary instructions for miscellaneous functions, storing and retrieving NMR data and shim sets, adjustment of the integral trace, VNMR keyboard commands, parameter file and macro listings, and checkout requirements.
| Operating frequency | 300 MHz |
|---|---|
| Probe (Inova-300-1) | 5mm 4-Nucleus Plus probe |
| Probe (Inova-300-2) | 5mm Autoswitchable probe |
| Minimum sample volume | 0.7 mL |
| TMS concentration | Not more than 0.1% |
| Default 1H scans | 8 |
| Default 13C scans | 256 |
Sample volume should be at least 0.7 mL.
Do not use more than 0.1% TMS.
Click the ABORT ACQ button, which aborts acquisition and discards data.
The default is nt=8.