3M Attest Biological Indicator Quality Control for Table-top Steam Sterilizers

3M has launched a product called TM AttestTM, which is a sterile U-shaped network tutorial for quality control of table-top steam sterilizers. The product complies with the AAMI Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities (ANSI/AAMI ST79:2006/A1:2008/A2:2009)1, and is applicable to all healthcare facilities that perform steam sterilization, including clinics and dental offices that use table-top sterilizers.


3M Attest Biological Indicator Quality Control for Table-top Steam Sterilizers - cover page
Brand
3M
Category
Health Care
Document type
User Manual
Language
English
Pages
7
File format
PDF
File size
2.3 MB
Published
24 November, 2023
Updated
24 November, 2023
MD5 checksum
742196642B2F8CF2D8CFCC48AB53D4B4

About this manual

This document provides quality control procedures for table-top steam sterilizers using 3M Attest Biological Indicators. It covers monitoring tools including physical, chemical, and biological indicators, with specific guidance on BI selection based on sterilizer type and cycle temperature. The guide also explains how to prepare and use biological indicator test packs and interpret results.

It includes frequently asked questions about BI testing frequency, incubation requirements, and compatibility with different sterilizer models and cycles, with reference to AAMI ST79 standards.

What's inside

  • Introduction
  • What is a table-top steam sterilizer?
  • How do they work?
  • Manufacturers
  • Frequently Asked Questions
  • Physical Monitors
  • Chemical Indicators
  • Biological Indicators
  • BI Testing Procedure
  • BI Selection by Cycle Type

Specifications

BI Type for Gravity 270°F3M™ Attest™ 1261 Biological Indicator
BI Type for Gravity and Prevacuum 250°F and 270°F3M™ Attest™ 1262* Biological Indicator
BI Type for Gravity 270°F Rapid Readout3M™ Attest™ 1291** Rapid Readout Biological Indicator
BI Type for Gravity 250°F Rapid Readout3M™ Attest™ 1292** Rapid Readout Biological Indicator
BI Type for Prevacuum 270°F Rapid Readout3M™ Attest™ 1292** Rapid Readout Biological Indicator

Frequently asked questions

What monitoring tools should be used for table-top steam sterilization cycles?

All steam sterilizers used for reprocessing of patient care items should be routinely monitored with physical monitors, chemical indicators (CIs), biological indicators (BIs), and process challenge devices (PCDs).

How is routine BI testing conducted in a table-top steam sterilizer?

Biological indicator testing is conducted in a fully loaded chamber with the BI PCD placed in the most challenging location, typically the 'cold point' or least favorable area for steam penetration.

Which 3M™ Attest™ Biological Indicator should be used to monitor a table-top steam sterilizer?

The appropriate 3M Attest Biological Indicator depends on the sterilizer’s air-removal method and cycle temperature. For example, 3M™ Attest™ 1262 is universal for all gravity-displacement cycles.

How often do I need to incubate a positive control biological indicator?

A positive control BI should be incubated every day a test BI is run, and if multiple test BIs from the same lot are run, only one control BI from that lot needs to be incubated.