3M has launched a product called TM AttestTM, which is a sterile U-shaped network tutorial for quality control of table-top steam sterilizers. The product complies with the AAMI Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities (ANSI/AAMI ST79:2006/A1:2008/A2:2009)1, and is applicable to all healthcare facilities that perform steam sterilization, including clinics and dental offices that use table-top sterilizers.
This document provides quality control procedures for table-top steam sterilizers using 3M Attest Biological Indicators. It covers monitoring tools including physical, chemical, and biological indicators, with specific guidance on BI selection based on sterilizer type and cycle temperature. The guide also explains how to prepare and use biological indicator test packs and interpret results.
It includes frequently asked questions about BI testing frequency, incubation requirements, and compatibility with different sterilizer models and cycles, with reference to AAMI ST79 standards.
| BI Type for Gravity 270°F | 3M™ Attest™ 1261 Biological Indicator |
|---|---|
| BI Type for Gravity and Prevacuum 250°F and 270°F | 3M™ Attest™ 1262* Biological Indicator |
| BI Type for Gravity 270°F Rapid Readout | 3M™ Attest™ 1291** Rapid Readout Biological Indicator |
| BI Type for Gravity 250°F Rapid Readout | 3M™ Attest™ 1292** Rapid Readout Biological Indicator |
| BI Type for Prevacuum 270°F Rapid Readout | 3M™ Attest™ 1292** Rapid Readout Biological Indicator |
All steam sterilizers used for reprocessing of patient care items should be routinely monitored with physical monitors, chemical indicators (CIs), biological indicators (BIs), and process challenge devices (PCDs).
Biological indicator testing is conducted in a fully loaded chamber with the BI PCD placed in the most challenging location, typically the 'cold point' or least favorable area for steam penetration.
The appropriate 3M Attest Biological Indicator depends on the sterilizer’s air-removal method and cycle temperature. For example, 3M™ Attest™ 1262 is universal for all gravity-displacement cycles.
A positive control BI should be incubated every day a test BI is run, and if multiple test BIs from the same lot are run, only one control BI from that lot needs to be incubated.