The FLUKE 323/324/325 Clamp Meter Calibration Manual provides instructions on how to calibrate the FLUKE 323, 324, and 325 clamp meters.
This document serves as the official calibration manual for the Fluke 323, 324, and 325 clamp meters. It provides essential safety information, detailed electrical and mechanical specifications, and lists the required equipment for calibration. The guide is designed for technicians to perform accurate performance tests and adjustments on these specific models.
| AC Current Range (323) | 400.0 A |
|---|---|
| AC Current Range (324, 325) | (40.00, 400.0) A |
| AC Voltage Range | 600.0 V |
| DC Voltage Range | 600.0 V |
| Resistance Range (323, 324) | (400.0, 4000) Ω |
| Resistance Range (325) | (400.0, 4000, 40000) Ω |
| Operating Temperature | -10 °C to +50 °C |
| Battery Type | 2 AAA |
The manual lists specific calibrator outputs for various functions, such as 1.05 V for AC Volts and 400 Ω for Ohms.
After adjusting all calibration points, the display shows 'Save'; push the HOLD button to save the data.
The accuracy is 2.0 % ±5 digits for frequencies between 45 - 65 Hz.