AIRVO 2 User Manual, intended for healthcare professionals, describes the features and how to use the product in detail.
This manual is intended for healthcare professionals and describes reprocessing procedures for the Fisher & Paykel 900PT600 Disinfection Kit used with the AIRVO 2 / myAIRVO 2 humidifier.
It covers cleaning, high-level disinfection using the supplied disinfection tube and filter, and correct storage between patients.
It is used to achieve high-level disinfection of the AIRVO 2 / myAIRVO 2 humidifier outlet elbow.
Do not clean the left-hand chamber port, because this can damage the non-return valve located at that port.
The AIRVO 2 humidifier must be cleaned and disinfected between patients, including cleaning, high-level disinfection, and storage.
The listed items include mild detergent, 70% alcohol solution or wipes, and clean disposable lint-free cloths.