Endress+Hauser Liquiphant FailSafe FTL80/81/85 with Nivotester FailSafe FTL825 Functional Safety Manual

Liquiphant FailSafe with Nivotester FailSafe is a safety-rated level measuring system for liquids.


Endress+Hauser Liquiphant FailSafe FTL80/81/85 with Nivotester FailSafe FTL825 Functional Safety Manual - cover page
Brand
Endress+Hauser
Category
Monitoring Equipment
Document type
User Manual
Language
English
Pages
40
File format
PDF
File size
1.6 MB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
515019DAE0249C300DA54FE33797575A

About this manual

This document is a functional safety manual for the Liquiphant FailSafe FTL80/81/85 level limit measuring system. It covers Version I with Nivotester FailSafe FTL825 and Version II using only the Liquiphant FailSafe sensor for minimum or maximum level monitoring of liquids in safety systems according to IEC 61508 Edition 2.0.

It describes system structure, permitted device types, safety requirements and boundary conditions, installation, operation, device configuration, maintenance, proof testing, repairs, return, commissioning or proof test protocols, and the certificate.

What's inside

  • Structure of the measuring system
  • Permitted device types
  • Safety function
  • Restrictions for use in safety-related applications
  • Information on Version I
  • Information on Version II
  • Installation
  • Operation
  • Device configuration
  • Proof test

Specifications

Safety functionPoint level detection (minimum or maximum)
Version I outputTwo safety contacts (terminals 13/14 and 23/24)
Version II output4 to 20 mA analog signal
MIN mode OK current18.5 mA
MAX mode OK current13.5 mA
MIN mode demand current9.0 mA
MAX mode demand current6.0 mA
Fault alarm current< 3.6 mA
Short-circuit alarm current> 21.0 mA
LIVE signal0.25 Hz square wave, ±0.5 mA amplitude
Valid firmware versionas of 01.00.00
Valid hardware versionas of 01.00 (electronics)
SIL capabilityup to SIL 3 according to IEC 61508 Edition 2.0

Frequently asked questions

What is the safety function of the measuring system?

The measuring system's safety function is point level detection, implemented as maximum or minimum point level detection.

Which two versions of the measuring system are described?

Version I uses Liquiphant FailSafe with Nivotester FailSafe, and Version II uses only Liquiphant FailSafe.

How does the output signal behave in MIN mode when the fork is covered?

It reports status OK with a nominal current of 18.5 mA, including the LIVE signal.

What proof test information is included in the manual?

The manual includes proof test procedures and commissioning or proof test protocols for Version I and Version II.