Nikon Eclipse Ti2 Microscope User Manual

The inverted Nikon Eclipse Ti2 microscope, equipped with a high precision encoded stage and Perfect Focus System is specifically designed for automated acquisition of large image files (time lapse, stitched images, high-content screening), and fast ratiometric imaging. Furthermore, a laser engine makes the system capable of TIRF and FRET imaging.


Nikon Eclipse Ti2 Microscope User Manual - cover page
Brand
Nikon
Category
microscope
Document type
User Manual
Language
English
Pages
29
File format
PDF
File size
11.8 MB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
9C65393E7CE673FDFADD738F5A5487B9

About this manual

This document is a user guide for the Nikon Eclipse Ti2 inverted microscope system, which is equipped with a high precision encoded stage and Perfect Focus System for automated acquisition of large image files, fast ratiometric imaging, and TIRF and FRET imaging with a laser engine.

It covers user guidelines, microscope general information and care, startup and sample setup, Köhler illumination, NIS Elements acquisition setup, basic and advanced image acquisition, image processing, macros, data management, and shutdown.

What's inside

  • Eclipse Ti2 specifications
  • User guidelines
  • Microscope general information and care
  • Startup and sample setup
  • System startup
  • Accessory equipment
  • Sample setup and observation with the microscope
  • Köhler Illumination
  • NIS Elements acquisition setup
  • Basic image acquisition

Specifications

CameraHamamatsu Orca Flash4.0 LT camera
LED excitation lines395, 440, 470, 508, 555, 640 nm
Fluorescence filtersDAPI, CFP, GFP/FITC, YFP, RFP/TRITC, CY5
Fura-2 illuminator340 nm and 380 nm
Laser engine488 nm and 561 nm lasers
Objectives4x, 10x (Air); 20x, 40x (Air; ELWD, long working distance); 60x, 100x (Oil), 100x TIRF (Oil)
SoftwareNIS Elements with AR (Advanced Research) and HC (High Content) modules
StageHigh precision encoded stage
Perfect Focus SystemPerfect Focus System (PFS)
Immersion oilType F provided by the facility
Coverslip thickness (oil objectives)0.17 (1.5) µm

Frequently asked questions

What immersion oil should be used with the Nikon Eclipse Ti2?

Use only the immersion oil (Type F) provided by the facility and do not mix it with other substances.

What supplies are users expected to bring?

Users must bring their own supplies, including pipets, slides, coverslips, and computer disks; the facility provides immersion oil and lens paper.

How should the lenses be cleaned after use?

After use, clean the lenses with lens paper only, repeating until no immersion oil remains on the paper.

How is Köhler illumination set?

Focus the image, close and centre the field diaphragm, adjust condenser height and aperture diaphragm, then open the field diaphragm until its edge disappears.