The Nikon user manual provides safety instructions, setup steps, and operating guidelines for the Eclipse Ti‑E, Ti‑E/B inverted microscopes and the…
About this manual
This manual explains safe handling, installation, and operation of Nikon's inverted microscopes. It includes detailed safety symbols, cautions for hot components, biohazard handling, and guidelines for using motorized accessories such as the Perfect Focus Unit.
The document also lists technical specifications, power requirements, and recommended environmental conditions for optimal performance, along with assembly instructions and maintenance procedures.
What's inside
- Introduction
- Safety Precautions
- LED Safety
- Power Supply and AC Adapter
- Motorized Nosepiece and Hub Controller
- Handling Hazardous Specimens
- Assembly and Installation
- Specifications
Specifications
| Dimensions (W x D x H) | 260 mm × 559 mm × 729 mm |
|---|
| Mass | 26.5 kg |
|---|
| Objectives | CFI60 |
|---|
| Eyepieces field number | 22 |
|---|
| Nosepiece sockets | Six |
|---|
| Focusing stroke | 10 mm |
|---|
| Coarse focus per rotation | 300 µm |
|---|
| Fine focus per rotation | 50 µm |
|---|
| Lamp rating | 12 V 100 W halogen |
|---|
| Specified lamp model | OSRAM HLX 64623 or PHILIPS 7724I |
|---|
| Average lamp life | 2000 hours |
|---|
| Power supply input rating | 100‑240 VAC, 1.8 A, 50‑60 Hz |
|---|
| Operating temperature range | 0 °C to 40 °C |
|---|
| Maximum relative humidity | 85 % RH (no condensation) |
Frequently asked questions
What safety steps must be taken when replacing the lamp?
Wait about thirty minutes after turning off the lamp, ensure the lamp and lamphouse have cooled, turn off the power switch, unplug the power cord, and wear gloves or use protective clothing.
What power source does the microscope use?
The microscope is powered by an AC adapter that operates with 100 to 240 VAC at 50‑60 Hz.
How is the LED in the Perfect Focus Unit classified?
The LED is classified as a Class 1 LED product according to IEC/EN 60825‑1:2001.