SWP SWP7035-Ex Isolated Safety Barrier Instruction Manual

SWP7035-Ex Isolated Safety Barrier Instruction Manual introduces the features and specifications of the product.


SWP SWP7035-Ex Isolated Safety Barrier Instruction Manual - cover page
Brand
swp
Category
Other device
Document type
User Manual
Language
Chinese
Pages
2
File format
PDF
File size
819 KB
Published
16 November, 2023
Updated
16 November, 2023
MD5 checksum
DCA2779516D1F5CD4A2D666779C69A09

About this manual

The SWP SWP7035-Ex is an analog input isolated safety barrier. It supplies a fully floating DC power source to a two-wire 4/20 mA transmitter in a hazardous area and transmits the transmitter's 4/20 mA current signal, isolated, from the hazardous side to the safe side to drive a safe-area load.

The manual describes precautions, main technical parameters, operating environment, product certification, intrinsic safety loop matching, wiring, dimensions, and installation and removal. It also lists safety parameters, cable distribution parameters, and mounting on a 35 mm standard rail. This manual is written in Chinese.

What's inside

  • Precautions
  • Overview
  • Main technical parameters
  • Operating environment
  • Product certification
  • System composition
  • Wiring
  • Dimensions
  • Installation and removal

Specifications

ChannelsSingle input, dual output, fully floating
Transmitter locationZones 0, 1, 2; Groups II A/B/C; T4-T6
Safe-area output signal4-20 mA or 1-5 V
Safe-area load resistance≤550 Ω (current output); ≥10 kΩ (voltage output)
Hazardous-area input signal4-20 mA
Hazardous-area supply voltage>15.5 V at 20 mA; ≤26 V open circuit
Conversion accuracy0.1%
Temperature drift (-20 to 60°C)<0.1 μA/°C
Power indicatorGreen LED
Power supply20-35 Vdc
Power consumption24 V, 120 mA
Maximum voltage Um250 V
Explosion-proof marking[Ex ia Ga] II C
Safety parameters (terminals 9-10-11)Uo=28 V, Io=93 mA, Po=0.66 W; II C Co=0.083 μF, Lo=4.2 mH; II B Co=0.366 μF, Lo=12.6 mH; II A Co=1.016 μF, Lo=33.6 mH
Safety isolation voltage250 V ac
Insulation strength2500 V ac for 1 min
Continuous operating temperature-20 to +60°C
Storage temperature-40 to +80°C
Relative humidity10-90% RH
Atmospheric pressure80-110 kPa
Certification bodyNEPSI
Certification standardsGB3836.1-2010; GB3836.4-2010
Cable distribution capacitanceCk = 0.1154 μF/km
Cable distribution inductanceLk = 0.20 mH/km
Dimensions99 mm (W) × 114.5 mm (H) × 22.5 mm (D)
Weight140 g
Terminal cable size0.5-1.5 mm² single or multi-strand
Mounting rail35 mm standard rail; conforms to 35×27×7.3 DIN46277
Warranty18 months from shipment

Frequently asked questions

What power supply voltage does the SWP7035-Ex use?

20 to 35 Vdc.

What output signals are available on the safe side?

4-20 mA or 1-5 V.

What are the dimensions of the SWP7035-Ex?

99 mm wide by 114.5 mm high by 22.5 mm thick.

What is the explosion-proof marking of the SWP7035-Ex?

[Ex ia Ga] II C.

Download PDF (819 KB)

More swp Manuals